background image

Troubleshooting

September 2018

TRB-UM-00478-EN-01

Page 10 

TROUBLESHOOTING

The typical failure mode for a turbine flow meter is failure to provide an output indication 
of flow  Troubleshooting this condition can be divided into two major sections: mechanical 
and electrical 

Mechanical Issues

Mechanical items to check when there is a failure-to-indicate condition:

•  Is there flow present to the instrument? 

Confirm all isolation and/or bypass valves are in the correct orientation to provide a 

moving fluid stream to the unit  Spectacle blinds and other blocking devices must be 

removed in order to allow fluid to enter the meter and turn the rotor  Flange gaskets 

can also slip out of place during installation and block the flow stream 

•  Is there foreign material blocking the rotor? 

Debris or solids allowed to enter the flow meter can easily stop the rotation of the 

rotor  Purging lines for construction debris should be performed with the instrument 

removed from the line  Remove the meter and remove all foreign bodies  Typical debris 

is Teflon thread tape, flux, rags and slug from the bottom of tanks  Anytime upstream 

pipe maintenance is performed, there is a potential of bearing contamination or 

blockage 

•  Is there damage to the internal assembly? 

Allowing solids to enter the meter or shocking the system with extreme pressurization 

variations can damage the internal assembly such that the rotor will no longer rotate  

Remove the meter and gently blow air through the meter  Verify that the rotor can 

freely move  Do not use high pressure air or further damage can result  If the rotor does 

not rotate freely under minimal input, contact the factory for repair 

Electrical Issues

Electrical items to check when there is a failure-to-indicate condition:

•  Is the pickup or flow computer connected? 

The pickup or flow computer must be connected to the data acquisition system for 

signal to be received  Often, a break is encountered in the cable run rather than at the 

connector 

•  Is power supplied to the pickup or flow computer? 

Several models of pickup and all flow computers require supply power to operate  

Consult the applicable wiring diagram for your model and verify that supply voltage is 

present on the correct conductors at the correct pins 

•  Is the electrical installation correct? 

Many factors affect electrical installation including power and signal cable proximity, 

shielding from adjacent electrical noise sources, proper grounding, amplifier gain 

settings, length of cabling runs, impedance matching of pickup and amplifier or 

readout, shielding from excessive heat sources, and more  Each installation is unique 

and may require several iterations of troubleshooting to correct electrical issues  

Badger Meter can assist in solving any issues that may arise  Contact Technical Support 

Содержание Exact Dual Rotor

Страница 1: ...TRB UM 00478 EN 01 September 2018 User Manual Turbine Flow Meters Exact Dual Rotor...

Страница 2: ...alves and Piping 8 Calibration 10 Maintenance 11 Bearing Replacement 11 Visual Inspection 11 Recommended Spare and Replacement Parts 11 Troubleshooting 12 Mechanical Issues 12 Electrical Issues 12 Flo...

Страница 3: ...n of damage that might have occurred during shipment NOTE N If damage is found request an inspection by the carrier s agent within 48 hours of delivery and file a claim with the carrier A claim for eq...

Страница 4: ...rs reduces the fall off of the output frequency at low flow rates extending the usable range or turndown ratio Additionally swirl at the meter inlet which causes a significant inaccuracy in a single r...

Страница 5: ...ousing on a common stem The stem is threaded into the downstream support except for the 1 4 in 84 and 3 8 in 86 sizes where it is captured on both ends with a nut Stem nuts are held in place with high...

Страница 6: ...on the Swirl Effect The Cox Exact Dual Rotor uses two closely coupled rotors turning in opposite directions The flow exiting from the first rotor greatly affects the inlet incidence angle on the rear...

Страница 7: ...an deform the end connection of the flow meter resulting in sealing issues and problems in performing routine maintenance Stainless steel is relatively soft and can be damaged if not handled with care...

Страница 8: ...e changes encountered and provide an accurate linearized output MAKE SURE THE CALIBRATION OF THE FLOW METER MATCHES THE APPLICATION FLUID CONDITIONS IN ORDER TO ACHIEVE ACCURACY UNCERTAINTY AND REPEAT...

Страница 9: ...application Visual Inspection Simple visual inspection is the only other preventive maintenance to be performed Periodically inspect the flow meter housing for signs of damage wear or fatigue failure...

Страница 10: ...ing the system with extreme pressurization variations can damage the internal assembly such that the rotor will no longer rotate Remove the meter and gently blow air through the meter Verify that the...

Страница 11: ...0 2 32 0 600 310 228 60 118 110 2 27 1173 5 103 69 53 573 79 2 1 2 40 1 000 500 381 00 190 500 3 78 1892 7 172 82 86 409 35 3 48 2 000 800 762 00 304 800 7 57 3028 345 64 138 254 4 64 5 000 1500 1905...

Страница 12: ...2 10 3 8 8 6 0 50 12 70 2 75 69 85 2 42 61 47 0 87 22 10 1 2 8 0 50 12 70 2 75 69 85 2 42 61 47 0 87 22 10 5 8 10 0 625 15 88 2 92 74 17 2 42 61 47 1 00 25 40 3 4 12 0 75 19 05 3 25 82 55 2 53 64 26 1...

Страница 13: ...3 8 8 6 0 375 9 53 2 75 69 85 2 42 61 47 0 87 22 10 1 2 8 0 50 12 70 2 75 69 85 2 42 61 47 0 87 22 10 5 8 10 0 75 19 05 2 92 74 17 2 42 61 47 1 00 25 40 3 4 12 0 75 19 05 3 25 82 55 2 53 64 26 1 37 34...

Страница 14: ...25 133 3 5 25 133 3 1 1 2 24 1 50 38 10 6 00 152 4 2 80 71 12 5 00 127 0 6 12 155 4 6 12 155 4 2 32 2 00 50 80 6 50 165 1 4 30 109 2 6 00 152 4 6 50 165 1 6 50 165 1 2 1 2 40 2 50 63 50 7 00 177 8 4 3...

Страница 15: ...ction 20 24 1 1 2 End Connection 24 32 2 End Connection 32 40 2 1 2 Available for flanges only 40 48 3 Available for flanges only 48 64 4 Available for flanges only 64 End ConnectionType 37 MS Flare A...

Страница 16: ...Badger Meter Europe PO Box 341442 Dubai Silicon Oasis Head Quarter Building Wing C Office C209 Dubai UAE 971 4 371 2503 Slovakia Badger Meter Slovakia s r o Racianska 109 B 831 02 Bratislava Slovakia...

Отзывы: