Te
chnical S
p
ecifications
SurgiStat Low-Power Electrosurgical Generator Service Manual
3-5
Performance Characteristics
The system presents audible and visible alarms when it does not sense a patient return
electrode.
•
When a fault condition occurs, the alarm indicator flashes red, an alarm tone sounds,
and the system disables output power.
•
The red LED alarm indicator remains illuminated until the condition that caused the
alarm is corrected.
•
Activation attempts during an alarm condition result in an audio alarm and the alarm
indicator flashes.
•
When the alarm condition is resolved, the green single or split-plate indicator will
illuminate.
•
The system measures the return electrode sensing current according to IEC 60601-1.
Low Frequency (50–60 Hz) Leakage Current
High Frequency (RF) Leakage Current
Split-Plate
Set resistance: 10
Ω
± 5
Ω
to 135
Ω
± 10
Ω
Continuous measurement:
Once the system establishes the split-plate electrode
resistance, an increase of 40% in resistance will cause
an alarm. When the alarm condition exists, the system
deactivates output power.
Enclosure source
current, ground open
< 300 μA
Source current, patient
leads, all outputs
Normal polarity, intact ground: < 10 μA
Normal polarity, ground open: < 50 μA
Reverse polarity, ground open: < 50 μA
Sink current at high
line, all inputs
< 20 μA
Bipolar RF leakage
current
< 39 mA
rms
Monopolar RF leakage
current
(additional tolerance)
< 150 mA
rms
Содержание SurgiStat
Страница 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Страница 2: ......
Страница 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Страница 10: ......
Страница 34: ......
Страница 54: ......
Страница 74: ......
Страница 88: ......
Страница 100: ...9 12 SurgiStat Low Power Electrosurgical Generator Service Manual Connector Panel Block Connector Panel Block ...
Страница 110: ......
Страница 111: ......