Surgistat Low-Power Electrosurgical Generator Service Manual
Rear Panel
Rear Panel
Layout of connectors and controls on the rear panel
①
Equipotential Connector
Allows attaching a standard grounding cable to chassis ground for additional
protection against leakage current.
②
Volume Control
Controls the volume of the audible tones produced during normal unit activation. To
increase volume, rotate the knob clockwise.
③
Footswitch Receptacle
Accepts the E6008 or E6008B monopolar footswitch. Use the monopolar footswitch
for both monopolar and bipolar activation. Use only a Covidien monopolar footswitch
with the SurgiStat Low-Power Electrosurgical Generator. Use of an incompatible
footswitch may cause unexpected output.
④
Power Cable Receptacle and Fuse Holder
Connects a hospital grade power cord to supply AC mains power to the unit.
Avertissement: Risque du feu.
Remplacez les fusibles
comme marquès.
T3.15AL250V
WARNING: Risk of Fire.
Replace Fuse as Marked.
T3.15AL250V