7-8
SurgiStat Low-Power Electrosurgical Generator Service Manual
Correcting Common Problems
Blank or confusing LED
display
Faulty ribbon cable between
main board and display board
Contact your Covidien Service
Center.
Display board malfunction
Mode buttons do not
operate correctly when
pressed
Loose or disconnected cable
between main board and display
board
Contact your Covidien Service
Center.
Loose or disconnected cable
between front panel overlay and
display board
Damaged front panel overlay
Generator is on and
the accessory is
activated, but
generator does not
deliver output
Malfunctioning footswitch or
handswitching instrument
1) Turn off the generator. Check
and correct all accessory
connections.
2) Turn on the generator.
3) Replace the accessory if it
continues to malfunction.
Display board malfunction
Contact your Covidien Service
Center.
Power set too low
Increase the power setting.
An error condition exists
1) Check the Cut and Coag
displays for an error code
number.
2) Note the number and refer to
the error codes descriptions in
this section.
Main board malfunction
Contact your Covidien Service
Center.
RF output stage malfunction
Footswitch will not
activate output
Malfunctioning or damaged
footswitch receptacle
Contact your Covidien Service
Center.
Footswitch activation signal lost
on main board
Incompatible footswitch
Use only a Covidien footswitch with
the SurgiStat generator.
Situation
Possible cause
Recommended action
Содержание SurgiStat
Страница 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Страница 2: ......
Страница 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Страница 10: ......
Страница 34: ......
Страница 54: ......
Страница 74: ......
Страница 88: ......
Страница 100: ...9 12 SurgiStat Low Power Electrosurgical Generator Service Manual Connector Panel Block Connector Panel Block ...
Страница 110: ......
Страница 111: ......