5
Sensor Saturation Accuracy
Low Perfusion
70 to 100% ± 2 digits
Adult
70 to 100% ± 2 digits
Adult and Neonate with Motion
70 to 100% ± 3 digits
*as tested with PM1000N monitor
Refer to each instrument operator’s manual for clinical testing summary and clinical
study Bland-Altman plots.
Optical Specifications
LED wavelength (Approximate)
Output power
Red: 650 to 670 nm
3.0 mW
Infrared: 880 to 910 nm
4.0 mW
Wavelength range can be especially useful to clinicians.
Environmental Specifications
Operational temperature range
0°C to 50°C
Storage temperature range
-40°C to 70°C
Operational and Storage Relative humidity range
15% to 95% non condensing
Additional Copies of Instructions
Additional copies of these Instructions are available at no charge by calling Covidien
or its authorized distributors. Also, permission is hereby granted under Covidien
copyrights to purchasers of products obtained from Covidien or its authorized
distributors to make additional copies of these instructions for use by such purchasers.
Any other use of this sensor is not authorized by Covidien under any patents.