EN-12
Kendall SCD 700 Series
Alarm Code
Alarm Type
Description
Troubleshooting
Temperature Alarm
Manual reset required
If the internal case
temperature of the
Controller drops below
5°C (41°F) or exceeds 55°C
(131°F).
High temperature: Make
sure the controller is not
covered by bedding and
that the fan port, located
near the power cord is not
obstructed.
Low Temperature: Allow
the system to warm to
room temperature
E9
E9
Battery Alarm
Service Required
Safe battery operation of
the controller can not be
ensured.
Service Technician
Only: Ensure that an
unauthorized battery
pack replacement has not
been made. Replace pack
or return to Covidien for
service.
E10
Tubing Disconnect Alarm
User Resettable
Pressure measured in
the inflatable garment is
below 10 mmHg for 10
consecutive cycles or no
garments are detected
during startup.
Check for disconnected
tube sets or garments and
reconnect.
A
B
E12
Pressure Transducer Alarm
Service Required
The system could not sense
a pressure rise of more
than 5 mmHg during an
inflation cycle or during
start up.
Service Technician Only:
Check the transducer tube
inside the controller and
ensure it is neither kinked
or disconnected.
E13
Low Battery Alarm
Recharge Battery
There is less than 15
minutes of battery charge
remaining. The pump and
valves will continue to
operate for as long as there
is enough power.
Plug the Controller into an
AC power outlet.