EN-10
Kendall SCD 700 Series
Alarm Codes
Alarm Code
Alarm Type
Description
Troubleshooting
Garment Mismatch Error
User Resettable
The Garment Detection
procedure has detected
a garment configuration
(Leg or Foot flashing green)
that does not match the
User-selected configuration
(Leg or Foot red).
Press the port
configuration button(s)
to turn the foot selection
on/off depending on
what type of garment(s)
is connected to the
Controller. If the proper
garment is selected and
the problem persists have
the Controller serviced by a
professional.
A
B
System High Pressure Alarm Manual Reset required
System pressure has
exceeded 90 mmHg (Leg
sleeve) or 180 mmHg (Foot
Cuff).
Check for kinked tubes or
patient interference with
the garments, like pressing
foot against foot board
A
B
E1
High Pressure (Leg Sleeves)
User Resettable
Leg Sleeve pressure is
greater than 47 mmHg
for 10 consecutive cycles;
or pressure is above 65
mmHg for 5 consecutive
cycles.
Check for a tight leg
sleeve and adjust fit
appropriately. Also check
for a partially occluded
tube.
A
B
E2
High Pressure (Foot Cuffs)
User Resettable
Foot Cuff pressure is
greater than 135 mmHg
for 10 consecutive cycles
or pressure is above 160
mmHg for 5 consecutive
cycles.
Check for a tight foot
cuff and adjust fit
appropriately. Also check
for a partially occluded
tube.
A
B
E2
Low Pressure (Leg Sleeves)
User Resettable
Leg Sleeve pressure is
less than 43 mmHg for 10
consecutive cycles.
Check for a loose leg
sleeve and adjust fit
appropriately. Also check
for leaks in the sleeve or
the tube connections
A
B
E3
Low Pressure (Foot Cuffs)
User Resettable
Foot Cuff pressure is less
than 125 mmHg after 5
consecutive cycles.
Check for a loose foot
cuff and adjust fit
appropriately. Also check
for leaks in the cuff or the
tube connections
A
B
E3