Footswitch Receptacles
2-10
Force FX Electrosurgical Generator C User’s Guide
Footswitch Receptacles
The rear panel contains three footswitch receptacles: two for monopolar and one for
bipolar.
Monopolar Footswitch Receptacles
You must connect a monopolar footswitch if you connect a monopolar footswitching
instrument to the generator. Use only a Valleylab monopolar footswitch with the Force
FX-C generator. Use of an incompatible footswitch may cause unexpected output.
Connect a two-pedal Valleylab monopolar footswitch to the Monopolar 1 Footswitch
receptacle.
The connected footswitch activates monopolar output for the instrument that is
connected to the Monopolar 1/CEM Instrument receptacle on the front panel.
Connect a two-pedal monopolar footswitch to the Monopolar 2 Footswitch receptacle.
The connected footswitch activates monopolar output for the instrument that is
connected to the Monopolar 2 Instrument receptacle on the front panel.
Bipolar Footswitch Receptacle
You must connect a bipolar footswitch if you connect a bipolar footswitching instrument
to the generator.
Connect a single-pedal bipolar footswitch to the Bipolar Footswitch receptacle.
The connected footswitch activates bipolar output for the instrument that is connected
to the Bipolar Instrument receptacle on the front panel.