Setting Up the Generator
4-4
Force FX Electrosurgical Generator C User’s Guide
4. Plug the generator power cord into a grounded receptacle.
5. Turn on the generator by pressing the power switch on ( | ). Verify the following:
–
All visual indicators and displays on the front panel illuminate.
–
Activation tones sound to verify that the speaker is working properly.
6.
If the self-test is successful
, a tone sounds. Verify the following:
–
Indicators above the default mode buttons (Standard bipolar, Pure cut, and
Fulgurate coag) illuminate green.
–
Each display shows a power setting of 1 watt.
–
The REM Alarm indicator illuminates red.
If the self-test is not successful
,
an alarm tone sounds. A number may momentarily
appear in the Cut display and, in most cases, the generator is disabled. Note the
number and refer to
Once the self-test is successful, connect the accessories and set the generator controls.
Refer to
Preparing for Bipolar or Macrobipolar Surgery
Preparing for Ultrasonic Electrosurgery
13.
Important
If the coag mode has been optionally changed to default to Desiccate or Spray, that
corresponding indicator illuminates after the self-test is performed successfully.