Output Characteristics
Technical Specifications
Force FX Electrosurgical Generator C User’s Guide
9-15
CEM Coag
Output Waveforms
Instant Response technology, an automatic adjustment, controls all bipolar modes and all
cut modes. It does not control the coag modes because of their fulguration capabilities.
As tissue resistance increases from zero, the generator outputs constant current followed
by constant power followed by constant voltage. The maximum output voltage is
controlled to reduce capacitive coupling and video interference and to minimize sparking.
Bipolar
Monopolar Cut
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
45
50
55
60
65
70
Precise
470 kHz sinusoid
Standard
470 kHz sinusoid
Macro
470 kHz sinusoid
Low
390 kHz sinusoid. Similar to the Pure cut mode except
the maximum voltage is limited to a lower value.
Pure
390 kHz sinusoid
Blend
390 kHz bursts of sinusoid, recurring at 27 kHz
intervals; 50% duty cycle