17
4.1.2
Using the hand pendant control
WARNING
The rear casters of the patient lift should not be locked when lifting an individual.
Locked casters could cause the lift to tip and endanger the patient and assistants. It
is recommended that the rear casters be left unlocked during lifting procedures to
allow the patient lift to stabilize itself when the patient is initially lifted from a chair,
bed or any stationary object.
To raise the patient lift, press
○
1
the boom
UP button ( up arrow) to raise the boom
and the patient.
To lower the patient lift, press
○
2
the boom
DOWN button ( down arrow) to lower
the boom and the patient.
○
3
alarms the user when the load exceeds
the lift capacity, and the indicator light will
come on.
○
4
displays battery status with light
indicators.
Standard Lifter
Bariatric Lifter
Where the hand pendant is absent or at fault, the control box can be used to lower
and raise the patient lift.
The buttons for raising and lowering the lift is located on the front/top of the control
box.
Содержание L400C
Страница 15: ...14 Standard Lift Control Box Bariatric Lift Control Box ...
Страница 37: ...36 APPENDIX SERVICE RECORD DATE PERFORMED BY CONDITION REPORT ...
Страница 38: ...37 DATE PERFORMED BY CONDITION REPORT ...
Страница 39: ...www costcaremed com 1 714 349 8331 ...