CA 8010 Instruction Manual
- 9 -
Calibration
Once you have made the settings you require, and when no key is pressed for a short
time the display shows „SH 99“. After the heating-up phase of the sensor the display
shows „BLOW“. You are now ready to begin with gassing the sensor.
1.
Run:
In the first step you gasing the sensor with a calibration solution of 0,20‰.
Please confirm the measuring result which is indicated by pressing the OK
button.
2.
Run:
Start the second run by pressing the OK button. The display shows again
„SH 99“, the sensor heats up and the display shows again “BLOW”. Now you
start with the second gassing by using a calibration solution of 1,00‰. Please
also confirm the second measuring result and finalise the calibration
procedure by pressing the OK button.
After the completetion of this works the CA 8010 is ready for operation for the next
6 month.
Order information – for the CA 8010
* VAT to be added to all prices plus shipping costs, if necessary. Subject to change.
Product
Number
Item no.
Price*
Breathalyzer CA 8010 black
Shipment in a aluminium case, with accessories comprising:
25 mouthpieces, batteries, belt pouch, instruction manual
1. piece
8010
Mouth pieces with rebreathing inhibitor
1 VE = 25 St. einzeln verpackt
1. piece
0023
Belt pouch
For the convenient storage of the device and mouth pieces
1. piece
7202
Transport- and storage case
1. piece
USB jack
For the connection of the Alco Conect 8010 to a PC or Laptop
1. piece
Gas simulator exclusiv
Complete with thermostat for constant temp. 34°C, air pump for
calibration, flow meter with control and automatic air-flow stop.
1 piece
0403
Calibration-alcohol-solution
for Cosmos Gas simulators 500 ml with 0,20 ‰ + 1,00 ‰
1 bottle
0404