CA 8010   Operation 

Inserting the rechargeable battery (Akku) 

Open  the  battery  cover  at  the  backside  of  the  CA  8010.  Insert  the  rechargeable 
battery. This is also the place of the buffer-battery for the CA 8010. The guarantee 
expires if the label removed. 
The backup (buffer) battery ensures that the functionality of the memory and time is 
maintained in the case of the rechargeable battery runs low. 

 
 
 
 
 
 
 

Loading the rechargeable battery  

Please connect the CA 8010 to the supplied USB cable and a free USB port on your 
PC,  Notebook  or  car  adapter  for  loading.  The  battery-status-  indication  on  the 
display shows the loading procedure by the three moving bars.  If the bars comes to 
a standstill the rechargeable battery is fully charged. 
 

Switch on the device 

Briefly press the power button to switch on the CA 8010. This is confirmed by an 
audible signal and the display shows the current time. 
 

 
 
 
 
 
 
 
 
 

Buffer battery 

Содержание CA 8010

Страница 1: ...2 WORD A4 E 10 10 C A 8010 GA Kal GA Alle Sprachen...

Страница 2: ...ion the AlcoConnect 8010 7 Switch on the device 8 Calibration mode 8 Settings 8 Country code 8 Measuring unit 8 Alert level 8 Calibration 9 Order information for the CA 8010 9 Troubleshooting 10 For y...

Страница 3: ...measurements 3 Measuring result in 4 Simple and fast calibration can also be performed by the client himself To do that Cosmos provides 1 mobile devices Gas simulator For further information please a...

Страница 4: ...lead to criminal culpability In Germany criminal prosecution and a fine can be expected if neurological impairment is 0 30 or above A test with the CA 8010 will not be accepted by the police as eviden...

Страница 5: ...s maintained in the case of the rechargeable battery runs low Loading the rechargeable battery Please connect the CA 8010 to the supplied USB cable and a free USB port on your PC Notebook or car adapt...

Страница 6: ...tes before a test 1 Switch on the CA 8010 by press briefly the power button 2 By pressing the OK button for more than one second the display shows PH Now the sensor is pre hated 3 Please place the mou...

Страница 7: ...to the time 199 measuring results will be stored if the memory is full the oldest result is automatically deleted Time and date adjustment Please press the lateral down key for 3 seconds The second in...

Страница 8: ...00 with the number 20 by pressing the key and confirm with OK Now the third and fourth 00 flashes Replace these two 00 with the number 10 by pressing the key Please do not press the OK button or any o...

Страница 9: ...essing the OK button After the completetion of this works the CA 8010 is ready for operation for the next 6 month Order information for the CA 8010 VAT to be added to all prices plus shipping costs if...

Страница 10: ...ries do not throw into fire do not charge batteries Risk of explosion do not use force to open Risk of explosion Only dispose of as toxic waste Comply with local regulations governing disposal For you...

Страница 11: ...l sensor fuel cell Precision 0 05 at 0 50 Measurement range 0 10 to 4 00 0 05 to 2 00 mg L Temperature range 5 C bis 50 C Shutdown automatic after 30 seconds The CA 8010 should be recalibrated every 6...

Страница 12: ...de without authorisation from us Applicable EC guidelines Emissions Radiated emission EN55022 Interference immunity Electrostatic discharge EN61326 1 Electromagnetic field EN61326 1 Magnetic field pow...

Отзывы: