background image

CA 8010 Instruction Manual 

 
 

 

 

                          

         - 8 - 

Calibration the AlcoConnect 8010 

 

Switch on the device 

Press briefly the power button to switch on the CA 8010. This is confirmed by an 
audible signal and the display shows the current time. 
 
 

Calibration mode 

Aktivate the calibration mode by pressing the central lateral  keys between 6 and 7  
+ OK-button (10) at the same time. The display shows „

0 0 0 0

“, and the first two 

digits start flashing.  Replace the two flashing 

00 

with the number  

20

 by pressing 

the  „+“ key and confirm with 

OK.

 Now the third and fourth 

00

 flashes. Replace 

these two 

00 

with 

the number 

10 

 by pressing the “+” key. 

!

 Please 

do not

 press the OK button or any 

other key. After a few seconds the device switches automatically into the calibration 
mode.   
 

Settings 

Factory pre-set for germany (21) in ‰ and alert 0,50 ‰ 

 

Country code: 

You can change the country code by pressing the „

+

“ button. 

 
Number 20   

 

Switzerland 

Number 21   

   Germany  

Number 23   

 

England 

 
Measuring unit

You can change the measuring unit by pressing the „OK“ button.. 
 


‰ 
mg/l  

 

Alert level: 

You can change the alert level by pressing the „-„ button. 
 

0,20‰ 
0,30‰ 
0,50‰ 
0,80‰ 

 

Содержание CA 8010

Страница 1: ...2 WORD A4 E 10 10 C A 8010 GA Kal GA Alle Sprachen...

Страница 2: ...ion the AlcoConnect 8010 7 Switch on the device 8 Calibration mode 8 Settings 8 Country code 8 Measuring unit 8 Alert level 8 Calibration 9 Order information for the CA 8010 9 Troubleshooting 10 For y...

Страница 3: ...measurements 3 Measuring result in 4 Simple and fast calibration can also be performed by the client himself To do that Cosmos provides 1 mobile devices Gas simulator For further information please a...

Страница 4: ...lead to criminal culpability In Germany criminal prosecution and a fine can be expected if neurological impairment is 0 30 or above A test with the CA 8010 will not be accepted by the police as eviden...

Страница 5: ...s maintained in the case of the rechargeable battery runs low Loading the rechargeable battery Please connect the CA 8010 to the supplied USB cable and a free USB port on your PC Notebook or car adapt...

Страница 6: ...tes before a test 1 Switch on the CA 8010 by press briefly the power button 2 By pressing the OK button for more than one second the display shows PH Now the sensor is pre hated 3 Please place the mou...

Страница 7: ...to the time 199 measuring results will be stored if the memory is full the oldest result is automatically deleted Time and date adjustment Please press the lateral down key for 3 seconds The second in...

Страница 8: ...00 with the number 20 by pressing the key and confirm with OK Now the third and fourth 00 flashes Replace these two 00 with the number 10 by pressing the key Please do not press the OK button or any o...

Страница 9: ...essing the OK button After the completetion of this works the CA 8010 is ready for operation for the next 6 month Order information for the CA 8010 VAT to be added to all prices plus shipping costs if...

Страница 10: ...ries do not throw into fire do not charge batteries Risk of explosion do not use force to open Risk of explosion Only dispose of as toxic waste Comply with local regulations governing disposal For you...

Страница 11: ...l sensor fuel cell Precision 0 05 at 0 50 Measurement range 0 10 to 4 00 0 05 to 2 00 mg L Temperature range 5 C bis 50 C Shutdown automatic after 30 seconds The CA 8010 should be recalibrated every 6...

Страница 12: ...de without authorisation from us Applicable EC guidelines Emissions Radiated emission EN55022 Interference immunity Electrostatic discharge EN61326 1 Electromagnetic field EN61326 1 Magnetic field pow...

Отзывы: