76
11 INTERFACE
•
HH and LL ratio setting
Curtailed command
WRTO_VV<CR>
Regular command
#SS_00_WRTO_VV:GG<CR>
Data field
Symbol
Data type
Unit Min. Max. Remarks
Identification No.
S
2 digit integer
00
09
Command
WRTO ASCII code
Data
V
2 digit integer
%
10
90 Can be set every 10% from 10 to 90%.
Checksum
G
2 digit integer
00
FF Data from “#” to “:” is summed up and its
two’s complement is taken.
•
Leak rate limit setting (HI limit and LO limit)
Curtailed command
Command_±VVVV.VVV<CR>
Regular command
#SS_00_CC_Command_±VVVV.VVV:GG<CR>
Data field
Symbol
Data type
Unit
Min.
Max.
Remarks
Identification No.
S
2 digit integer
00
99
Channel #
C
2 digit integer
00
15
Command
ASCII code
Data
V
7 digit
fixed point
Leak
rate unit ±0000.000 ±0999.000
Checksum
G
2 digit integer
00
FF
Data from “#” to “:” is
summed up and its two’s
complement is taken.
•
Timer and Equivalent internal volume or leak coefficient setting
Curtailed command
Command_VVVV.VVV<CR>
Regular command
#SS_00_CC_Command_VVVV.VVV:GG<CR>
Data field
Symbol
Data type
Unit
Min.
Max.
Remarks
Identification No.
S
2 digit integer
00
99
Channel #
C
2 digit integer
00
15
Command
ASCII code
Data
V
Fixed point
0000.000
9999.000
Checksum
G
2 digit integer
00
FF
Data from “#” to “:” is
summed up and its two’s
complement is taken.
Example:
Writing command “WDET_0000.500” programs DET timer 0.5s.
Limit ratio can be set every 10% from 10 to 90% of the leak rate limits (Hi and
Lo limits). The tester transmits error if other numbers are received.
NOTE
Setting will be ineffective, and tester transmits NAK code every 0.1 seconds
Δ
P
exceeds ±1000 Pa while mL/min is selected for leak rate unit.
NOTE
Set the END timer 0.2 seconds and DL1, DL2, BAL and DET timers 0.1 second
or longer. Values are rounded off to one decimal place. Value larger than
999.8 will be infinite.
NOTE
Writing “WBLO_-0500.000” sets the leak rate limit –500 Pa.
NOTE
Setting will be ineffective, and tester transmits NAK code
Δ
P exceeds
±1000 Pa while mL/min is selected for leak rate unit.
NOTE
When memory switch 24 is set to 0, the maximum values for CHG
(pressurization) time and STB (stabilization) time are 9999.000. If the time
value exceeds 1000.000, the digits after the decimal point are truncated. If the
value exceeds 6000.000, it is set as infinite with no timeout. In this case, the
maximum value is forced to 9999.000.
NOTE
Содержание LS-1866
Страница 2: ......
Страница 6: ......
Страница 22: ......
Страница 34: ......
Страница 56: ......
Страница 62: ......
Страница 71: ...11 INTERFACE 69 g Typical PLC connection NPN h Typical connection with PLC PNP ...
Страница 88: ......
Страница 89: ...13 APPENDIX 87 1 13 3 APPENDIX 1 13 3 1 1 E EX XT TE ER RN NA AL L A AP PP PE EA AR RA AN NC CE E ...