background image

MOBILTEX® DATA LTD. 

Calgary, Alberta 

www.corTalk.com 

TITLE:  

uDL2 Configuration and Operation Guide

 

DOCUMENT NO.: 

UDL2-MAN-001

 

SHEET: 

22 of 28

 

REV: 

1.07 

 

5

 

Operation 

There are no user controls on the actual uDL2.  Data acquisition cannot be scheduled to start at a particular date and 
time.  Instead, the device automatically powers up when the measurement cable is installed and immediately starts data 
acquisition activities.  The device ceases data acquisition and powers down when the measurement cable is removed.  

The waterproof connector that mates to the uDL2 has an automatic physical locking function. The white arrow on the 
cable should align with the black arrow on the top graphic overlay of the uDL2. When the cable is inserted, it will “click” 
to indicate that the connector is locked in place. New cables may require application of additional force until the “click” 
is heard in order to properly position the connector internal sealing o-ring.  The cable is released by pulling back the 
black ring (which has the white arrow) of the connector.   

The  measurement  cable  has  three  banana  plugs  to  allow  connection  directly  to  banana  jacks.  There  are  also  three 
extension clip wires included that can attach to the banana plugs of the measurement cable.  The unit can be operated in 
two modes: coupon mode and dual potential mode.   

Coupon Mode

 

The black wire is connected to the coupon, the blue wire is connected to the structure (pipe), and the red wire is connected 
to the reference half-cell.  

When connected to a coupon test station, ensure that the test station coupon disconnect 

switch is left in the “open” or “off” position; otherwise, current will not flow through the uDL2 internal current 
sense.

 Figure 28 shows typical attachments to a coupon test station.  The optional W164I0ULD03 cable allows direct 

attachment to the DE-9P test port connector on the CC Technologies Coupon Test Station. 

    

V

A

Coupon

Black

Pipe
Blue

Ref Cell

Red

uDL2 (Coupon Mode)

 

Figure 28 Sample Cable Attachment to a Coupon Test Station 

 

 

 

 

 

 

Содержание uDL2

Страница 1: ...D Calgary Alberta www corTalk com TITLE uDL2 Configuration and Operation Guide DOCUMENT NO UDL2 MAN 001 SHEET 1 of 28 REV 1 07 uDL2 Data Logger for CP Coupons Configuration and Operation Guide Print Date March 26 2018 ...

Страница 2: ...Revision 1 07 Name Location Job Function Signature Date Tony da Costa Calgary Engineering Manager Jerry Chilibecki Calgary Director of Engineering REVISION HISTORY Rev Start Date Approval Date Description Prepared By Draft 0 01 Feb 19 2013 First draft Tony da Costa 1 00 Feb 20 2013 Initial release Tony da Costa 1 01 April 9 2013 Added coupon test station connection picture Added CC Tech coupon tes...

Страница 3: ...Config From uDL2 13 4 3 1 2 Write Config To uDL2 13 4 3 1 3 Link Status 13 4 3 1 4 Factory Options 13 4 3 1 5 Firmware 14 4 3 1 6 Serial 14 4 3 1 7 Read Write Config File 14 4 3 1 8 Set uDL2 Clock 14 4 3 1 9 Read Faults 14 4 3 1 10 Detailed Status 14 4 3 1 11 Reboot uDL2 15 4 3 2 Configuration Group 15 4 3 2 1 Timing 15 4 3 2 2 Coupon Potential Sampling CH1 or Potential Sampling CH1 16 4 3 2 3 Cou...

Страница 4: ...n Application Installation Dialog 11 Figure 15 uDL2 Configuration Application Installation Complete 11 Figure 16 Driver Installation Screen 12 Figure 17 uDL2CONFIG Main Screen 13 Figure 18 Faults Dialog 14 Figure 19 Detailed Status Dialog 15 Figure 20 Channel 2 Mode Selection 16 Figure 21 Coupon Area Converter 16 Figure 22 Overcurrent Detector Retries Configuration Dialog 17 Figure 23 Extract To M...

Страница 5: ... potential mode where the potential of two structures may be measured relative to a common reference point A large LCD can be used to locally verify the readings Measurement storage capacity of 2 000 000 readings and rechargeable battery allow the device to capture readings for over 4 months An internal GPS receiver provides an accurate time reference and site geo location coordinates The small ph...

Страница 6: ...ct manual for further ratings information before making connections to the product Do not use an input marked with measurement category I for measurements within measurement categories II III or IV Do not connect category II inputs directly to mains a Class 2 CSA UL transformer must be used for isolation Do Not Operate Without Covers Do not operate this product with covers or panels removed Avoid ...

Страница 7: ...t Contents The Mobiltex part number for the uDL2 kit is A20A03UDL20 non USA or A20A03UDL21 USA The following items are included in each kit uDL2 Data Logger for CP Coupons Unit A15000UDL20 non USA A15000UDL21 USA Figure 3 uDL2 Front View uDL2 USB Cable W164IFUDL01 Figure 4 uDL2 USB Cable uDL Black Fabric Case H20712UDL01 Figure 5 uDL Black Fabric Case ...

Страница 8: ...lips H27400112BK H27400112RD H27400112BU Figure 7 Large Alligator Clips Programming Application and Driver Media CD or USB flash drive Figure 8 Programming Application and Driver USB Flash Drive To use a PC to communicate with the uDL2 you will require the media CD or USB flash drive that shipped with the uDL2 The media contains drivers and the uDL2 configuration application If you do not have the...

Страница 9: ... Interface Cable Mobiltex P N W164I0UDL03 Figure 9 CC Technologies Coupon Test Station Direct Interface Cable USA Canada AC USB Wall Charger P11606UDL01 Figure 10 USA Canada AC USB Wall Charger DC USB Vehicle Charger P11606UDL02 Figure 11 DC USB Vehicle Charger Test Station Spacer Rings Provide More Height Inside Standard Test Station H23510RMU01 Black H23530RMU01 Orange H23540RMU01 Yellow H23550R...

Страница 10: ...he application software and drivers require that the PC be running Windows XP SP3 or higher for an operating system The PC must have an available USB port 4 2 Software Installation To install the configuration application and driver software insert the media CD or USB flash drive that came with the configuration interface into the appropriate drive or USB port on the PC If autorun is enabled on th...

Страница 11: ...in the installation screens clicking Next to move to the next screen each time Some screens may pause for tens of seconds as the Windows installation procedure is processing The final screen should show the dialog in Figure 15 Click Close to complete the installation Figure 15 uDL2 Configuration Application Installation Complete A shortcut to the configuration application and the uDL2 manual are c...

Страница 12: ...he following screen should display Follow the installation instructions presented by the driver installer application Figure 16 Driver Installation Screen Once the driver installer completes you may then plug in the uDL2 through the USB cable into an available USB port on the PC 4 2 3 Adobe Acrobat Reader XI Adobe Acrobat Reader XI is provided on the installation media to allow viewing of the PDF ...

Страница 13: ...g or not connected an error dialog will pop up If it is not connected simply connect the USB cable to the computer and try again If the uDL2 LCD display is showing bSY wait until USb is shown before attempting to read the configuration from the uDL2 Once communications have been established with the uDL2 the operational parameters can be configured for the requirements of a particular measurement ...

Страница 14: ...led in the configuration the uDL2 clock should be set at least once before data acquisition is attempted Beyond that the clock should be updated periodically to resynchronize the clock to actual time The internal clock of the uDL2 model can be expected to drift as much as 10s each month without GPS synchronization The uDL2 will maintain the real time clock as long as the battery is not fully deple...

Страница 15: ...ep mode to conserve battery power The Disconnect Cycle Period determines how often the coupon interruption relay will be opened for instant off and de polarization measurements The Disconnect Duration value is how long the relay is left open for each disconnect event After each relay transition the uDL2 will wait before taking the next DC potential measurement this wait time is determined by one o...

Страница 16: ...the uDL2 G2 G3 ranges Select the range based on the expected maximum coupon current and the range specifications from Appendix A Note that only the uDL2 G2 and uGL2 G3 models supports all three ranges earlier versions of the hardware only support the medium and high ranges The low range radio button will be disabled for units that do not support that range The Inv Polarity checkbox can be set if i...

Страница 17: ...attached and for 5 seconds after the analog cable is removed After the analog cable is attached the display will cycle through the measurement data and the time for 60 seconds Time is shown in 12 hour format with an A in the lower right of the display indicating AM times To enable this feature check the Enabled box The time may be displayed in UTC or local time Click on the corresponding radio but...

Страница 18: ... GMT time or in local time Local time is based on the time zone selected in the operating system of the computer being used to extract the data When the Extract To Master Data File button is clicked the following save as screen will prompt the user for the storage filename Figure 23 Extract To Master Data File Dialog The stored file is saved in a comma separated value CSV format that can be import...

Страница 19: ...MOBILTEX DATA LTD Calgary Alberta www corTalk com TITLE uDL2 Configuration and Operation Guide DOCUMENT NO UDL2 MAN 001 SHEET 19 of 28 REV 1 07 Figure 24 Sample Master Data File ...

Страница 20: ...ith processing applications are encountered For example Microsoft Excel 2003 has a limit of 65 536 rows while newer versions have a limit of 1 048 676 rows Enter the maximum number of lines rows desired in the output files and then click the Split button to break the input files into smaller files The output files will have the same names as the input files but suffixed with an incrementing number...

Страница 21: ...ed click on the Zoom All button When viewing DC Wave capture segments the currently displayed data segment may be selected by clicking the Previous Next First or Last buttons under the Wave Print Segment area These buttons do not appear for other preview chart types The current chart view may be saved to a file Click the Save Chart button to save a bitmap of the chart The bitmap may be saved in JP...

Страница 22: ...osition the connector internal sealing o ring The cable is released by pulling back the black ring which has the white arrow of the connector The measurement cable has three banana plugs to allow connection directly to banana jacks There are also three extension clip wires included that can attach to the banana plugs of the measurement cable The unit can be operated in two modes coupon mode and du...

Страница 23: ...y cases is a reference cell The red wire is connected to the first potential measurement point which may be a pipe or a structure The blue wire is connected to the second potential measurement point which again may be a pipe or a structure Figure 29 shows a typical attachment at a test station V1 V2 Ref Black Potential 2 Blue Potential 1 Red uDL2 Dual Potential Mode Figure 29 Dual Potential Analog...

Страница 24: ... digits of measurement information When the analog cable is first attached the uDL2 will enter a real time measurement display mode for a period of 1 minute During this time the uDL2 doesn t store information to memory and it provides real time measurement data on the LCD which is updated every 2 seconds The SAMPLE icon continuously illuminates during this real time display mode After the real tim...

Страница 25: ...f the over current disconnect is activated the relay will remain open until the analog is detached and then re inserted An over current disconnect is indicated by an LCD display showing O in the reading area To stop data logging simply unplug the measurement cable The device will automatically shut off Each time the measurement cable is inserted the device will begin the 1 minute real time measure...

Страница 26: ...lication and execute the uDL2 Firmware Update application from the Windows start menu The dialog in Figure 33 will be shown Figure 33 uDL2 Firmware Update Application Dialog To start the firmware update process first ensure that the uDL2 is connected to a USB port on the computer Next click on the Program Firmware button The application will indicate progress as the firmware is downloaded to the u...

Страница 27: ... latest uDL2 software package may be downloaded from the Mobiltex web site at the following URL http www mobiltex com cathodic udl2 uDL Registration Visit www mobiltex com service registration html and register your uDL to receive notification of free firmware updates Information provided will be used for the sole purpose of sending notification of firmware updates and will never be shared or sold...

Страница 28: ...red common Lightning Immunity Survives multiple 20KV 10mS surges Temp Measurement Accuracy 4 C 7 F over 20 to 60 C 4 to 140 F Potential DC Accuracy over 20 to 60 C 4 to 140 F 0 55 5mV Potential AC Accuracy over 20 to 60 C 4 to 140 F 1 5mV 20mV floor Current DC Accuracy over 20 to 60 C 4 to 140 F 0 5 2uA low range 0 5 20uA med range 0 5 750uA high range Current AC Accuracy over 20 to 60 C 4 to 140 ...

Отзывы: