4 | CORNING LAMBDA PLUS PIPETTORS
Nominal
Volume
(μL)
Cat. No.
Volume
(μL)
Accuracy
(%)
Precision
(%)
Fit to Tips
Without
Filter
(μL)
8-channel and 12-channel Pipettors
8 - 10
12 - 10
4080
4084
Min.
0.5
±10.0
≤8.0
10
5
±4.0
≤2.0
Max.
10
±2.0
≤1.2
8 - 50
12 - 50
4081
4085
Min.
5
±4.0
≤2.5
200
25
±3.0
≤1.2
Max.
50
±1.6
≤0.6
8 - 200
12 - 200
4082
4086
Min.
20
±3.0
≤1.5
200
100
±1.5
≤0.8
Max. 200
±1.0
≤0.6
8 - 300
12 - 300
4083
4087
Min.
50
±1.6
≤1.5
300
150
±1.2
≤1.0
Max. 300
±1.0
≤0.6
◗
Accuracy is the difference between the volume
dispensed and the test volume selected.
◗
Precision is the degree to which repeated measurements
under unchanged conditions show the same results.
The accuracy and precision are obtained gravimetrically
using Corning® pipet tips, performing at least 10
measurements of distilled water at a reference temperature
of 20°C ± 1°C according to EN ISO 8655 standard.
2. Packaging
The following components are supplied with the pipettor:
◗
Instruction manual
◗
Calibration key
◗
Ejector length adjustment spacers (for single-channel
pipettors)
◗
Ejector cap (for single-channel pipettor models 2 and 10)
◗
Identification labels
◗
Lubricant (for single-channel pipettors)
◗
Color identification rings (for single-channel pipettors).