12   |   CORNING LAMBDA PLUS PIPETTORS 

If the average aspirated volume exceeds the permissible 

value, the following should be performed:

1

2

3

4

K

1. Remove the pipetting pushbutton.  

WARNING:

 The pipetting pushbutton consists of 2 parts: 

the lower volume-setting section (A2) and the upper 

button (A1). After removal of the pushbutton, both parts 

will separate.

2. Holding the volume setting knob to protect it against 

rotation, insert the calibration key (K) into the cuts of the 

calibration screw.

3. Turn the calibration key clockwise to reduce the aspirated 

volume, or counterclockwise to increase the volume. 

One full turn of the calibration key changes the pipettor 

aspiration volume by the amount given in the table.

4. Take out the calibration key and fix the pipetting 

pushbutton. The pipetting pushbutton should be fixed 

by placing the lower section on the arbor first (A2) and 

then the button (A1).

Determine the average aspirated volume. The average 

volume should be within the permissible range shown 

in the table. If the volume exceeds the values stated, the 

recalibration procedure should be repeated.
When pipetting liquids with physical properties that are 

considerably different from those of water, follow the 

instructions as described in Section 5.

Содержание Lambda Plus

Страница 1: ...Corning Lambda Plus Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Страница 2: ...ng the Volume 5 4 Aspiration and Dispensing Instructions 6 5 Pre rinsing 8 6 Dense and Viscous Liquids 8 7 Corning Pipet Tips 9 8 Recommendations 9 9 Recalibration 10 10 Troubleshooting 13 11 Cleaning...

Страница 3: ...ettors are produced in four volume ranges 0 5 to 10 L 5 to 50 L 20 to 200 L 50 to 300 L Model Volume Range L Application 2 10 Measurement and transfer of micro volumes DNA sequencing and enzyme assay...

Страница 4: ...jector D Single channel Pipettor Ejector The ejector length can be adjusted to accommodate all common brands of pipet tips When using narrow tubes it may be necessary to remove the tip ejector This is...

Страница 5: ...ng at least 10 measurements of distilled water at a reference temperature of 20 C 1 C according to EN ISO 8655 standard 8 channel and 12 channel Pipettor Ejector The 8 and 12 channel pipettor construc...

Страница 6: ...the test volume selected Precision is the degree to which repeated measurements under unchanged conditions show the same results The accuracy and precision are obtained gravimetrically using Corning...

Страница 7: ...nd red Red numbers represent decimal fractions for microvolume pipettors or milliliters for macrovolume pipettors An example for each of these pipettors is given on the next page Nominal Volume L Coun...

Страница 8: ...too far below the required value the process must be repeated Set the locking ring into the upper position following the volume adjustment thus locking the knob and preventing accidental volume chang...

Страница 9: ...to the recommended depth or if the pipetting pushbutton is rapidly released air may enter the disposable pipet tip NOTE Avoid touching the orifice of the pipet tip Dispensing 3 Place the end of the pi...

Страница 10: ...sion are based on pipetting distilled water Handling of liquids with physical qualities of density viscosity and surface tension which differ substantially from water may need to be gravimetrically ch...

Страница 11: ...ry and should remain constant during aspiration The pipettor should be held in a vertical position Change the pipet tip when volume setting is changed or when a different liquid is to be aspirated Cha...

Страница 12: ...ical Density of the liquid used is close to that of distilled water Balance with appropriate sensitivity has been used Conversion factor mg L has been taken into account The requirements shown in Sect...

Страница 13: ...0 33 8 50 12 50 5 50 5 4 8 5 2 1 67 8 200 12 200 20 200 20 19 4 20 6 6 30 8 300 12 300 50 300 50 49 2 50 8 10 00 Perform 5 aspirations weigh each one and calculate the average value of the aspiration...

Страница 14: ...o reduce the aspirated volume or counterclockwise to increase the volume One full turn of the calibration key changes the pipettor aspiration volume by the amount given in the table 4 Take out the cal...

Страница 15: ...ns The pipet tip is incorrectly pressed onto the pipettor shaft Press the pipettor firmly The tip is damaged or worn out due to extensive use Replace the tip with a new one The pipettor incorrectly as...

Страница 16: ...eps No lubricant on the sealing elements Remove the tip ejector Unscrew the shaft nut and remove the shaft piston assembly seal and O ring Rinse the removed parts in distilled water and dry thoroughly...

Страница 17: ...ried and cooled to room temperature It is recommended to Unscrew the shaft nut slightly prior to sterilization and re tighten after autoclaving Set the locking ring in lower unlocked position prior to...

Страница 18: ...ion key M Ejector cap N Ejector pushbutton P Identification ring S Locking ring Please contact Corning Customer Service to inquire about availability of spare parts Pipettor model and name of the part...

Страница 19: ...not arising from defects in original material or workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with the transportation carrier In the...

Страница 20: ...re for research use or general laboratory use only Not intended for use in diagnostic or therapeutic procedures Not for use in humans These products are not intended to mitigate the presence of microo...

Страница 21: ...wan t 886 2 2716 0338 EUROPE CSEurope corning com France t 0800 916 882 Germany t 0800 101 1153 The Netherlands t 20 655 79 28 United Kingdom t 0800 376 8660 All Other European Countries t 31 0 206 59...

Отзывы: