6 – Corning Lambda EliteTouch Pipettors 

INSTRUCTION MANUAL – 7

A.

  To enable volume selection, set the black adjustment 

knob to the upper position.

B.

  To set a certain value, turn the knob. The volume is 

indicated by the counter and its value consists of four 

digits that should be read from top to bottom. Digits 

on the bottom drum represent the scale division for 

the pipettor model. See scale intervals in Section 5. 

The volume range of the pipettor model is shown on 

the counter overlay. 

C. 

After the desired volume has been set, move the knob 

down to avoid an inadvertent change of the volume. 

Examples of counter indications

The white mark 
separates the 
integer part from 
the decimal part of 
the aspirated liquid 
volume in µL.

Volume Ranges

0.1-2 µL

5-50 µL

100-1000 µL

Set Volume (μL)

0.1 µL

5 µL

100 µL

To attain the maximum accuracy, the set volume must be 

approached from a higher value by diminishing counter 

readings. Before reaching the desired value, reduce the 

speed of turning the adjustment knob to not exceed this 

value. 

7. Operating Instructions 

Observing the following recommendations will ensure 

maximum possible accuracy and precision of liquid sampling.

•  During operation, the pipettor’s volume setting should be 

locked, the black adjustment knob should be in the lower 

position.

A

B

C

The accuracy and precision are obtained gravimetrically using 

Corning® pipet tips performing at least 10 measurements 

of distilled water at a temperature of 20°C ± 1°C according 

to the EN ISO 8655 standard.
The design of the pipettor enables the user to recalibrate it 

according to the information presented in Section 9.

6. Pipettor Operation

Attaching the Tips

•  Attach the tip corresponding to the model on the pipettor 

on the shaft (Section 5).

•  Position the pipettor vertically when putting on the tips.
•  Single-channel pipettors, when attaching the tip, press 

the pipet tip on firmly using a slight twisting motion. 

This will ensure an airtight seal.

•  Multi-channel pipettors, when attaching the tips on the 

shafts, press the pipettor against the tips in the rack box 

until the shafts retract approximately 1.5 mm into the 

manifold. The suspension system of the shafts ensures even 

and tight sealing of the pipet tips. The rocking movement 

does not have to be performed to seal the pipet tips tightly.

  

NOTE:

•  Do not attach the tips with a swinging movement, 

as this may damage the shaft or plunger. 

Observe this 

rule particularly with single-channel pipettors of low 

volume range.

•  Never draw liquids directly into the pipettor without 

the tip attached.

Volume Setting

NOTE

: The red calibration switch above the counter 

should remain in the lower position during 

operation. Otherwise, any change in volume will 

result in a loss of calibration of the pipettor.

Содержание Lambda EliteTouch

Страница 1: ...Corning Lambda EliteTouch Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Страница 2: ...6051 0 5 10 20 6052 2 20 50 6053 5 50 100 6054 10 100 200 6055 20 200 1000 6056 100 1000 8 channel and 12 channel Pipettors 8 10 12 10 6057 6061 0 5 10 8 50 12 50 6058 6062 5 50 8 200 12 200 6059 606...

Страница 3: ...ector bar in the multi channel pipettors Materials Used Handle parts Polypropylene PP Polycarbonate PC Polyacetal POM Stainless steel Shaft and piston area parts Polypropylene PP Polyvinylidene fluori...

Страница 4: ...e the pipettor as it may damage the pipettor The single use of the tip reduces the risk of contamination of samples Keep the pipettor clean avoiding the use of aggressive substances e g acetone for cl...

Страница 5: ...ower position A B C The accuracy and precision are obtained gravimetrically using Corning pipet tips performing at least 10 measurements of distilled water at a temperature of 20 C 1 C according to th...

Страница 6: ...unter operating mode switch the calibration switch located under the CAL sign on the counter overlay This switch should only be used when the pipettor is calibrated upper position marked as CAL on the...

Страница 7: ...film remains relatively constant in successive pipetting operations with the same tip this error can be eliminated by forming the film before transferring the first sample This is done by aspirating a...

Страница 8: ...holding the handle unscrew the remaining part of the shaft holder set from it Then proceed according to the pipettor models shown below See ISO 8655 for the full table of the value of the conversion f...

Страница 9: ...d in the case of pipettor 2 L also the Seal P If you need to clean other components Unscrew the Shaft Cap M Remove from the Plunger I and pipetting Spring N Remove the Shaft J from the Connecting Bush...

Страница 10: ...g Spring N from the shaft Remove the Shaft J from the Connecting Bush L If necessary remove the Seal P from the Plunger I Check the condition of components and clean as recommended Replace any damaged...

Страница 11: ...ny damaged parts Assemble the components in reverse order When inserting the shaft or the shaft holder set into the Connecting Bush make sure that they are inserted all the way to the end by gently ro...

Страница 12: ...ealing elements e g after repeated autoclaving procedures Sterilization Sterilization in an autoclave The pipettor can be sterilized in an autoclave at 121 C for 20 minutes Sterilization under other c...

Страница 13: ...Shaft Kit with seal and O ring 2 SP9811 1 J O Shaft kit with O ring 10 SP9812 1 20 SP9813 1 J Shaft 50 100 SP9814 1 200 SP9815 1 1000 SP9816 1 L Connecting Bush All SP9810 1 H Tip ejector 2 10 SP9817...

Страница 14: ...natural forces or other causes not arising from defects in the original material workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with t...

Страница 15: ...EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 20 655 79 28 f 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Coun...

Страница 16: ......

Страница 17: ...2019 Corning Incorporated All rights reserved 3 19 CLS AN 525 H EliteTouch IM Corning Incorporated Daniszewska 4 03 230 Warsaw www corning com lifesciences Made in Poland...

Отзывы: