background image

Many times when attempting to detect the fetal heart, the maternal vascular sounds are 

heard instead of (or in some cases, in addition to) the fetal sounds.  These maternal sounds 

can come from one of the major arteries, the placenta or the umbilical cord.  The maternal 

vascular sounds are typically higher in frequency at a lower rate.  The heart rate calculation 

will display either the maternal rate or the fetal rate, whichever portion of the signal is 

stronger.  

If the fetal heart sounds cannot be located using the procedure as described above, a 

second exam should be performed using another commercially available fetal monitor as a 

repeated test.

5 and 8 MHz Vascular

Peripheral arterial sounds are typically higher in frequency.  For the best sounds, angle 

the probe approximately 45 degrees from the skin surface over the general location of 

the vessel.  Slowly move the probe side to side and vary the angle of the probe until 

the vascular sounds are heard.  Changing the angle of the probe has an effect on the 

frequency of the sound.  The steeper the probe angle is, the higher the frequency of the 

sound.

Peripheral venous sounds are not typically periodic and vary greatly depending on patient 

movement and breathing.  These sounds are more like the wind at the ocean and vary in 

pitch as the patient moves or breathes.

Obtaining Doppler Signals

Caution

   •  Doppler examinations should be performed only by trained individuals.

For any Doppler examination, it is essential that an adequate supply of coupling gel is used 

to transmit the ultrasound energy from the probe to the surface of the skin.  Re-apply more 

gel if it starts to dry out or spread so thinly that an air gap occurs between the probe and 

the skin.  It is not necessary to cover the entire surface of the probe, only the probe face.  

Applying too much gel makes the unit difficult to clean and does not aid in the performance 

of the probe.

Volume Slider

The audio level can be adjusted using the Volume Slider.  Moving the slider up will increase 

the volume, while moving it down will decrease it.

Signal Quality Indicator (TR250 only)

An inadequate signal can produce erroneous rates from the heart rate calculation.  The 

signal level that is being obtained is shown on the Signal Quality Indicator bars.  This 

indicator provides a visual aid in obtaining a strong audio signal by showing the pulsatile 

nature of the signal.  A large difference between the highest and lowest signal bars that are 

lit confirms that the quality of the signal is good and thus ensures the heart rate calculation 

is operating at peak performance.

The heart rate can be verified manually by counting the audible beats for 20 seconds and 

multiplying by 3, or for 15 seconds and then multiplying by 4.  Counting for less than 15 

seconds is not recommended due to a decrease in accuracy with the small sample size.

Obstetrical 

Fetal heart sounds are quite different from peripheral vascular blood flow sounds.  Fetal 

sounds are typically much lower in frequency and much higher in rate.  For early term fetal 

detection, start the probe at the pubic bone and slowly move along the midline – rocking 

the probe slowly from side to side until a heart beat is heard.  For mid to late term fetal 

detection the best chance of finding the heart sounds are to start on the fundus and move 

toward the navel and from one side of the abdomen to the other, slowly rocking the probe 

until the heart beat is heard.  The fetal heart reminds many people of a galloping horse and 

can vary in tone from a distant swishing sound to a hard clopping sound depending on the 

position of the baby and probe.

9

8

Содержание CTP-2

Страница 1: ...EUROPE gracht 20 he Hague rlands e USA Fetal and Vascular Dopplers User Manual for TR150 Series Units Non Display and TR250 Series Units Display man e ...

Страница 2: ...ccessories 13 Specifications 14 Warranty and Servicing Policy 17 Explanation of Symbols Back Cover Please read the manual carefully and become familiar with the operation features and maintenance of your Doppler prior to using the device or accessories Warra This wa warran Parts specifie for the repaire Labor 1 yea Center purcha Limitat indicat shall no by Coo Cooper terms s but not shippin or neg...

Страница 3: ...e questions or concerns regarding these warnings or contraindications please do not hesitate to contact CooperSurgical for further clarification Caution Dropping the TRIA II probe or accessories may cause damage to the housing or electronics Safety of Ultrasound The Tria II was designed with physician and patient safety in mind In early design phases all potential hazards were eliminated or reduce...

Страница 4: ...cture of the fetus or determine the fetal gender without a medical indication is inappropriate and contrary to responsible medical practice Although there are no confirmed biological effects on patients caused by exposures from present diagnostic ultrasound instruments the possibility exists that such biological effects may be identified in the future Thus ultrasound should be used in a prudent ma...

Страница 5: ... to ensure the highest quality standards SSQ Superior Sound Quality Every TRIA II Doppler is designed with a state of the art sound system that produces excellent sound quality and long term reliability LCD Display The LCD display TR250 only allows you to view the fetal heart rate in large easy to read digits monitor battery life observe signal strength indicators and view multiple diagnostic indi...

Страница 6: ...the 2 MHz probe are available 3 MHz Early and general purpose obstetrical examination This probe frequency is a general use model ideal for most stages of fetal examination Fetal heart sounds can be heard as early as 12 weeks and sometimes sooner depending on the position and size of the fetus 5 MHz This standard pencil style probe is an excellent vascular tool for locating deep specific vessels i...

Страница 7: ...al g rtment 5 Operation and Installation On Off Button Probe Clip Doppler Cable Connector Head Phone Connector Battery Compartment Door Belt Clip Volume Slider ...

Страница 8: ...at is being used This display will not reappear unless the probe is changed or the power is cycled in which case the display will again temporarily confirm the frequency of probe that is connected The unit then checks for proper internal operating temperature battery voltage reference voltage and power supply voltage levels If any of these characteristics are out of range the display will show the...

Страница 9: ...ge the batteries soon after the current examination is complete Non display units TR150 use the On Off LED as a battery indicator by flashing at a low rate approximately once per second when the battery level is low Several exams can still be performed in this state of operation The battery indicator will flash at a higher rate approximately twice per second when the battery level is very low The ...

Страница 10: ...visual aid in obtaining a strong audio signal by showing the pulsatile nature of the signal A large difference between the highest and lowest signal bars that are lit confirms that the quality of the signal is good and thus ensures the heart rate calculation is operating at peak performance The heart rate can be verified manually by counting the audible beats for 20 seconds and multiplying by 3 or...

Страница 11: ...ailable fetal monitor as a repeated test 5 and 8 MHz Vascular Peripheral arterial sounds are typically higher in frequency For the best sounds angle the probe approximately 45 degrees from the skin surface over the general location of the vessel Slowly move the probe side to side and vary the angle of the probe until the vascular sounds are heard Changing the angle of the probe has an effect on th...

Страница 12: ...y close attention to clean the seams along the plastic lines at the probe face but do not allow water or spray to enter through the connectors or speaker grill To disinfect unit use commercially available spray or wipe disinfectants registered with the EPA Clorox Broad Spectrum Quaternary Disinfectant is the only disinfectant that is CooperSurgical approved for use with the TRIA II product Follow ...

Страница 13: ...ist in lifting the batteries out Replace the batteries by paying close attention to the polarity indicators on the battery and the polarity indicators on the battery holder in the compartment Positive aligns with positive button and negative aligns with negative spring Insert the battery such that the spring contacts are loaded first and then press the battery firmly into place After all three hav...

Страница 14: ...eaker may cause poor sound Device damage from dropping the TRIA II probes or accessories Heart Rate inaccurate Try to relocate the probe for a better signal refer to Signal Strength Indicator For OB ensure maternal sounds are not mixing with fetal sounds Ensure by manual counting that the rate is between 50 and 220 BPM Battery indicator flashing Consult the Battery Monitoring section replace batte...

Страница 15: ... Temperature too high 6 5 Volt Supply too high 3 Reference Voltage too low 7 Battery Voltage too low 4 Reference Voltage too high 8 Battery Voltage too high Reference materials for Obstetrical and Peripheral Vascular testing Handbook of Fetal Heart Rate Monitoring Julian T Parer 1997 Doppler Ultrasound and Its Use In Clinical Measurement Peter Atkinson and John P Woodcock 1982 Noninvasive Diagnosi...

Страница 16: ...60601 1 4 IEC 60601 2 37 EN55011 Dimensions h x w x l 140 x 70 x 35 mm Weight 320 grams Operating temperature 10 to 40 C Operating humidity 30 to 75 Transport Storage temperature 20 to 50 C Transport Storage humidity 5 to 90 non condensing Battery type and voltage 3 AA Alkaline 1 5 volt for non rechargeable units 3 AA NiMH 1 5 volt for rechargeable units as applicable Battery life Batteries provid...

Страница 17: ...3 Mpa 0 09 Wo mW 12 8 0 013 fc MHz 5 3 5 3 5 3 Zsp cm 0 85 0 85 0 85 X 6 cm 0 4 0 4 y 6 cm 0 6 0 6 Az cm 0 4 Ele cm 0 8 Beam Dimensions EBD Transducer Model TSP 8 Operating Mode Continuous Wave cw Application s Peripheral Vascular ACOUSTIC OUTPUT MI ISPTA 3 ISPPA 3 mW cm2 W cm2 Global Maximum Value 0 03 229 0 23 Associated Acoustic Parameter Pr 3 Mpa 0 09 Wo mW 13 9 0 014 fc MHz 8 0 8 0 8 0 Zsp cm...

Страница 18: ...ut Parameters are measured in water Derated values denoted by the subscript 3 take into account a conservative level of attenuation that would be encountered in the human body The derated intensity values I 3 are obtained from water values of intensity Iw at a depth of z calculated by I 3 exp 0 23 0 3 f z Iw where f is the probe frequency in MHz and z is the depth in centimeters The derated peak r...

Страница 19: ...uct under the terms stated above and does not cover loss or damage from external causes such as but not limited to weather theft or abuse This warranty does not apply to batteries shipping case or other accessories or to damage from shipping tampering misuse or negligence Preventive maintenance the refinishing or replacement of any cosmetic defect or deterioration or the replacement of batteries u...

Страница 20: ...ered trademark of The Clorox Company Reorder Number Consult instructions for use U S Federal law restricts this device to sale by or on the order of a physician Date of Manufacture Serial Number Not made with natural rubber latex Authorized Representative in the European Community Manufacturer In order to preserve protect and improve the quality of the environment protect human health and utilize ...

Отзывы: