Battery Monitoring 

All TRIA

®

 II Doppler units perform continuous battery monitoring and give a visual indication 

of battery level.  Display units use a multiple level battery shaped indicator that indicates 

the voltage level of the battery.  The battery outline will flash when the battery level is very 

low, indicating that the user should change the batteries soon after the current examination 

is complete.

Non-display units (TR150) use the On/Off LED as a battery indicator by flashing at a low 

rate (approximately once per second) when the battery level is low.  Several exams can 

still be performed in this state of operation.  The battery indicator will flash at a higher 

rate (approximately twice per second) when the battery level is very low.  The user should 

change the batteries soon after the current exam is complete

Maintenance and Cleaning

Warnings

  •   The TRIA

®

 II  is not designed for liquid immersion.  Do not soak the Doppler main unit 

or probes in liquids.  Use only spray or wipe  cleaners and disinfectants. 

Do not use 

products containing bleach.

  •   The TRIA

®

 II  is not designed for sterilization processes such as autoclaving, gamma  

radiation or hydrogen gas.

  •   The TRIA

®

 II  is not intended to be used on open skin.  If there is evidence of open 

wound contamination, disinfect the probe before using again as described below.

The TRIA

® 

II  Doppler requires very little maintenance.  However, it is important to the 

continuing function of the unit and the health of the patients that the unit is cleaned and 

examined regularly per the following guideline:

After every examination:

Excess gel should be wiped off prior to docking the probe.  Probes and main unit should 

be cleaned with a damp cloth using warm water or presaturated isopropyl alcohol wipes 

or spray.  In particular, pay close attention to clean the seams along the plastic lines at the 

probe face but 

do not allow water or spray to enter through the connectors or speaker 

grill.

  

To disinfect unit, use commercially available spray or wipe disinfectants registered with 
the EPA.  Clorox

®

 Broad Spectrum Quaternary Disinfectant is the only disinfectant that is 

CooperSurgical approved for use with the TRIA

®

 II  product.  Follow the manufacturer’s 

instructions and wipe unit until it is dry of solutions.  Examiners should wash hands and 

change gloves after every exam.  Refer to local and hospital policies for cleaning and 

disinfection policies.

Store unit in a clean area free from dust and debris.  Follow temperature and humidity 

guidelines as specified at the end of this manual.

Warning

  •   If the unit is to be stored for longer than 90 days without use, remove the batteries prior 

to storage.  

Periodically (at least annually):

Inspect the main unit and probes for signs of cracks or breaks in the mechanical housing.  

Inspect cables and connectors for signs of wear or failure.  The user should discontinue use 

of the unit with any sign of loss of housing integrity.  Contact CooperSurgical for service.

It is recommended that rechargeable batteries be replaced annually.

7

10

Содержание CTP-2

Страница 1: ...EUROPE gracht 20 he Hague rlands e USA Fetal and Vascular Dopplers User Manual for TR150 Series Units Non Display and TR250 Series Units Display man e ...

Страница 2: ...ccessories 13 Specifications 14 Warranty and Servicing Policy 17 Explanation of Symbols Back Cover Please read the manual carefully and become familiar with the operation features and maintenance of your Doppler prior to using the device or accessories Warra This wa warran Parts specifie for the repaire Labor 1 yea Center purcha Limitat indicat shall no by Coo Cooper terms s but not shippin or neg...

Страница 3: ...e questions or concerns regarding these warnings or contraindications please do not hesitate to contact CooperSurgical for further clarification Caution Dropping the TRIA II probe or accessories may cause damage to the housing or electronics Safety of Ultrasound The Tria II was designed with physician and patient safety in mind In early design phases all potential hazards were eliminated or reduce...

Страница 4: ...cture of the fetus or determine the fetal gender without a medical indication is inappropriate and contrary to responsible medical practice Although there are no confirmed biological effects on patients caused by exposures from present diagnostic ultrasound instruments the possibility exists that such biological effects may be identified in the future Thus ultrasound should be used in a prudent ma...

Страница 5: ... to ensure the highest quality standards SSQ Superior Sound Quality Every TRIA II Doppler is designed with a state of the art sound system that produces excellent sound quality and long term reliability LCD Display The LCD display TR250 only allows you to view the fetal heart rate in large easy to read digits monitor battery life observe signal strength indicators and view multiple diagnostic indi...

Страница 6: ...the 2 MHz probe are available 3 MHz Early and general purpose obstetrical examination This probe frequency is a general use model ideal for most stages of fetal examination Fetal heart sounds can be heard as early as 12 weeks and sometimes sooner depending on the position and size of the fetus 5 MHz This standard pencil style probe is an excellent vascular tool for locating deep specific vessels i...

Страница 7: ...al g rtment 5 Operation and Installation On Off Button Probe Clip Doppler Cable Connector Head Phone Connector Battery Compartment Door Belt Clip Volume Slider ...

Страница 8: ...at is being used This display will not reappear unless the probe is changed or the power is cycled in which case the display will again temporarily confirm the frequency of probe that is connected The unit then checks for proper internal operating temperature battery voltage reference voltage and power supply voltage levels If any of these characteristics are out of range the display will show the...

Страница 9: ...ge the batteries soon after the current examination is complete Non display units TR150 use the On Off LED as a battery indicator by flashing at a low rate approximately once per second when the battery level is low Several exams can still be performed in this state of operation The battery indicator will flash at a higher rate approximately twice per second when the battery level is very low The ...

Страница 10: ...visual aid in obtaining a strong audio signal by showing the pulsatile nature of the signal A large difference between the highest and lowest signal bars that are lit confirms that the quality of the signal is good and thus ensures the heart rate calculation is operating at peak performance The heart rate can be verified manually by counting the audible beats for 20 seconds and multiplying by 3 or...

Страница 11: ...ailable fetal monitor as a repeated test 5 and 8 MHz Vascular Peripheral arterial sounds are typically higher in frequency For the best sounds angle the probe approximately 45 degrees from the skin surface over the general location of the vessel Slowly move the probe side to side and vary the angle of the probe until the vascular sounds are heard Changing the angle of the probe has an effect on th...

Страница 12: ...y close attention to clean the seams along the plastic lines at the probe face but do not allow water or spray to enter through the connectors or speaker grill To disinfect unit use commercially available spray or wipe disinfectants registered with the EPA Clorox Broad Spectrum Quaternary Disinfectant is the only disinfectant that is CooperSurgical approved for use with the TRIA II product Follow ...

Страница 13: ...ist in lifting the batteries out Replace the batteries by paying close attention to the polarity indicators on the battery and the polarity indicators on the battery holder in the compartment Positive aligns with positive button and negative aligns with negative spring Insert the battery such that the spring contacts are loaded first and then press the battery firmly into place After all three hav...

Страница 14: ...eaker may cause poor sound Device damage from dropping the TRIA II probes or accessories Heart Rate inaccurate Try to relocate the probe for a better signal refer to Signal Strength Indicator For OB ensure maternal sounds are not mixing with fetal sounds Ensure by manual counting that the rate is between 50 and 220 BPM Battery indicator flashing Consult the Battery Monitoring section replace batte...

Страница 15: ... Temperature too high 6 5 Volt Supply too high 3 Reference Voltage too low 7 Battery Voltage too low 4 Reference Voltage too high 8 Battery Voltage too high Reference materials for Obstetrical and Peripheral Vascular testing Handbook of Fetal Heart Rate Monitoring Julian T Parer 1997 Doppler Ultrasound and Its Use In Clinical Measurement Peter Atkinson and John P Woodcock 1982 Noninvasive Diagnosi...

Страница 16: ...60601 1 4 IEC 60601 2 37 EN55011 Dimensions h x w x l 140 x 70 x 35 mm Weight 320 grams Operating temperature 10 to 40 C Operating humidity 30 to 75 Transport Storage temperature 20 to 50 C Transport Storage humidity 5 to 90 non condensing Battery type and voltage 3 AA Alkaline 1 5 volt for non rechargeable units 3 AA NiMH 1 5 volt for rechargeable units as applicable Battery life Batteries provid...

Страница 17: ...3 Mpa 0 09 Wo mW 12 8 0 013 fc MHz 5 3 5 3 5 3 Zsp cm 0 85 0 85 0 85 X 6 cm 0 4 0 4 y 6 cm 0 6 0 6 Az cm 0 4 Ele cm 0 8 Beam Dimensions EBD Transducer Model TSP 8 Operating Mode Continuous Wave cw Application s Peripheral Vascular ACOUSTIC OUTPUT MI ISPTA 3 ISPPA 3 mW cm2 W cm2 Global Maximum Value 0 03 229 0 23 Associated Acoustic Parameter Pr 3 Mpa 0 09 Wo mW 13 9 0 014 fc MHz 8 0 8 0 8 0 Zsp cm...

Страница 18: ...ut Parameters are measured in water Derated values denoted by the subscript 3 take into account a conservative level of attenuation that would be encountered in the human body The derated intensity values I 3 are obtained from water values of intensity Iw at a depth of z calculated by I 3 exp 0 23 0 3 f z Iw where f is the probe frequency in MHz and z is the depth in centimeters The derated peak r...

Страница 19: ...uct under the terms stated above and does not cover loss or damage from external causes such as but not limited to weather theft or abuse This warranty does not apply to batteries shipping case or other accessories or to damage from shipping tampering misuse or negligence Preventive maintenance the refinishing or replacement of any cosmetic defect or deterioration or the replacement of batteries u...

Страница 20: ...ered trademark of The Clorox Company Reorder Number Consult instructions for use U S Federal law restricts this device to sale by or on the order of a physician Date of Manufacture Serial Number Not made with natural rubber latex Authorized Representative in the European Community Manufacturer In order to preserve protect and improve the quality of the environment protect human health and utilize ...

Отзывы: