Maintenance
52
6. Grasp the edge of the old lamp-unit and carefully pull it out of the
scanner.
7. Reinsert the new lamp-unit. Rest the power-connector side on the
slider and carefully push the lamp-unit into the scanner.
8. Insert the lamp-unit all the way so the internal power connections
engage with the connections on the lamp-unit.
9. Replace the lamp cover on the scanner - fit the bottom notches in the
holes and snap the cover shut.
The cover must always be closed
before turning scanner power ON.
Fit the cover
’
s bottom notches into the holes
10. Test the replacement by turning power and scanner on to see if the
lamp lights.
11. Reset the part replacement warning flag though WIDEsystem
’
s
“
Status
”
tab. Click on the lamp replacement message link to reset.
NOTE
: It is recommended to recalibrate after changing the lamp.