ECG1200G Electrocardiograph - User Manual
38
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
±
1
kV
for
input/output
lines
± 2 kV for power
supply lines
±
1
kV
for
input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1
kV
differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
dips,
short
interruptions
and
voltage
variations
on power supply
input lines
IEC 61000-4-11
< 5 % UT
(>95 % dip in
UT )
for 0,5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in
UT )
for 5 sec
< 5 % UT
(>95 % dip in UT )
for 0,5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the model ECG1200G
Electrocardiograph(ECG)
requires continued operation
during power mains
interruptions, it is recommended
that the model ECG1200G
Electrocardiograph(ECG) be
powered from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE UT is the a. c. mains voltage prior to application of the test level.