ECG1200G Electrocardiograph - User Manual
37
Appendix
EMC Information
Guidance and manufacturer’s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The model ECG1200G Electrocardiograph(ECG) is intended for use in the electromagnetic
environment specified below. The customer or the user of the model ECG1200G
Electrocardiograph(ECG) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
1
Guidance and manufacturer’s declaration – electromagnetic emission
2
The model ECG1200G Electrocardiograph(ECG) is intended for use in the electromagnetic
environment specified below. The customer or the user of the model
ECG1200GElectrocardiograph(ECG) should assure that it is used in such an environment.
3
Emissions test
Compliance
Electromagnetic environment - guidance
4
RF emissions
CISPR 11
Group
The model ECG1200G
Electrocardiograph(ECG) uses RF energy only
for its internal function. Therefor, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
5
RF emissions
CISPR 11
Class B
6
Harmonic
emissions
IEC 61000-3-2
Class A
7
Voltage
fluctuations /
flicker emissions
IEC 61000-3-3
Complies