ECG1200G Electrocardiograph - User Manual
29
Item
Options
Explanation
Print
Mode
[Auto 12*1]/[Auto 6*2+1]
/[Auto 6*2]/[Auto
3*4+2]/[Rhythm 12]
/[Rhythm 10 ]/[Rhythm 8]
/[Rhythm 6]
The selection will be used as the default printing
mode.
Lead
Gain
Smart/Current
The option selected will be used as printing gain
mode. "Smart" means the system will adjust gain
automatically to fit paper height; "Current" means
it will use screen waveform gain as that of
printing.
Auto
Strip
3Sec/4Sec/5Sec
/6Sec/8Sec/10Sec/
15Sec/20Sec/25Sec
The selection will be used as the time for printing
each strip.
Rhyth
m
Strip
10Sec/15Sec/20Sec
/25Sec/30Sec
In printing mode, when selecting rhythm 12,
rhythm 10, rhythm 8, or rhythm 6, the system will
use the time selected as the time for printing each
row waveform.
Avera
ge
QRS
[3*4+Mark]/[3*4]/
[None]
In printing mode, when selecting "Rhythm" Or
"Auto", the system will use the format selected to
print average QRS waveform.
Auto-
Diag
All/Data
/Conclusion/NonePrint
Printing diagnosis includes data and conclusion,
you can select as your need.
Perio
dic
[per1Min]/[per 2 Min]/[per3Min]
/[per5Min]/[per10Min]/[per20Mi
n]/[per30Min]/[per 60 Min]/[off]
When gathering ECG, the system will activate
printing according to the duration selected. When
printing mode is manual, it will print in "Auto
12*1" mode, or
in the mode currently set .
Print
devic
e
[Inside]/[Qutside A4]
choose prints out the ECG waveforms with
thermal printing system or USB external printer.
Print
depth
[depth]/[shallow]
Choose wave print depth.
Note: "Auto Strip", "Rhythm Strip", "Average QRS" , "Auto-Diag" , "Periodic" are only
available in auto and rhythm printing mode.
In [the Print Setup] interface, click [Adv-opr] to enter the print settings advanced
interface . as shown below: