5 Monitoring Contractions (External)

Symptom

Possible Cause

Solution

Worse   trace   quality   or 
fluctuant TOCO baseline

Too   tight   or   too   loose 
abdomen   belt   or   no 

elasticity

Ensure  the  abdomen  belt   has   been 
used   accurately   and   neither   too 

tight, nor too loose 

Maternal Movement

Relax patient’s spirit

Fetal Movement

Wait for a moment then monitor

Too high TOCO sensitivity 
(higher than 100 unit)

The body pressure from 
uterus   to   TOCO 

transducer   is   far   higher 
than the average value.

Insure favorable contact for patient 
skin with TOCO transducer. Change 

the position of TOCO transducer, if 
necessary.

Attachment 3 Monitoring Figure

Содержание CMS800G

Страница 1: ...CONTEC MEDICAL SYSTEMS CO LTD CMS800G FETAL MONITOR...

Страница 2: ...er and Removing Paper Jam 17 4 5 Power On the Monitor 17 4 6 Connect Transducers 17 5 Monitoring 17 5 1 Operation Procedure 17 5 2 Print Operation 20 5 3 Operation After Monitoring 21 6 Maintenance Ca...

Страница 3: ...nstallation of the relevant room complies with national standards and the instrument is used in accordance with the instructions for use Note This device is not intended for home use WARNING This devi...

Страница 4: ...sting The recommended testing interval is once per year The protection categories against electric shock of the patient connections are FHR1 FHR2 TOCO MARK This symbol indicates that the instrument is...

Страница 5: ...t a power cord with wet hands Make certain that your hands are clean and dry before touching a power cord WARNING The monitor should be installed by an authorized and qualified service engineer WARNIN...

Страница 6: ...sing the equipment immediately after moving it from a cold environment to a warm humid location CAUTION Do not immerse transducers in liquid When using solutions use sterile wipes to avoid pouring flu...

Страница 7: ...ring of two heart rates for twins This is achieved by using the facilities of two ultrasound transducers and an external contractions TOCO transducer with a recorder The monitor can display FHR TOCO M...

Страница 8: ...al broad beam ultrasound transducer is used for monitoring fetal heart rate FHR1 The ultrasound transducer operates at a frequency of 1 0MHz Put the ultrasound transducer on maternal abdomen to transm...

Страница 9: ...t Socket for FHR1 TOCO transducer and remote Marker FHR2 socket Socket for FHR2 Transducer Socket for Grounding Cable NET Socket for network Warning Symbol Push LCD Screen rotation lock 3 1 4 Main Int...

Страница 10: ...nnel it is I under single fetal monitoring mode which is default I II selectable under twins monitoring mode it can be changed through the main menu 3 Volume level ranging from 0 7 0 stands for sound...

Страница 11: ...setting parameters it contains alarm on off status alarm upper limit lower limit alarm postpone time print speed and print time 4 Indicate Bar In this item it includes system time bed No gestational a...

Страница 12: ...m when FHR is in alarm range When symbol appears the alarm indicator status is shut off Press the button to enable audio alarm the alarm indicator becomes when FHR is in alarm situation the alarm soun...

Страница 13: ...Function Freezing the screen Press the button to stop drawing and the screen becomes in frozen status press the button again to continue drawing This operation will clear the screen 8 Knob Key 1 Press...

Страница 14: ...orm Review 5 PRINT SET Print Setup Enter Print Setup 6 MONI TYPE Monitor Mode Optional single twins the default is single fetus 7 LANGUAGE Language Selection Optional Chinese CH English EN the default...

Страница 15: ...onds Note 1 When FHR is in alarm situation alarm indicator becomes red 2 When FHR exceeds the alarm limit and time exceed the set alarm delay time continuously alarm will occur and an alarm symbol wil...

Страница 16: ...ed patient information 14 SAVE Save Information Save related patient information return to previous menu 15 Exit Return to previous menu Note When adjustable range is letter or number numeral and lett...

Страница 17: ...r password under the item USR KEY before click CONFIRM This password is provided by manufacturer or distributor when Contec add new function to upgrade the system Version Revolving the knob key to ent...

Страница 18: ...r works properly please read this chapter and Chapter 1 Safety Guidance And follow the steps before using the monitor 4 1 Open the Package and Check Open the package and take out the monitor and acces...

Страница 19: ...r cabinet to check for a paper jam then feed the paper again Note Only use the manufacturer approved paper to avoid poor printing quality deflection or paper jam 4 5 Power on the Monitor WARNING If an...

Страница 20: ...nsducer to the belt Apply aquasonic coupling gel to the face of the transducer 2 Acquiring the Fetal Heart Signal Determine the location of the fetal heart using palpation or a fetoscope Place the ult...

Страница 21: ...the optimum position Positions with strong placental sounds swishing or fetal cord pulse indistinct pulse at fetal rate should be avoided If the fetus is in the cephalic position and the mother is sup...

Страница 22: ...al movement At the moment the mark will show in the correspond position of trend wave The count of fetal movement will add 1 after each push of the button Push the button and hold for one second then...

Страница 23: ...printing 5 3 Operation After Monitoring Remove transducers from patient Wipe transducer with a soft cloth to remove remaining ultrasound coupling gel Tear the paper at the folding place Switch off the...

Страница 24: ...rface of the monitor Use dry and soft cloth to wipe WARNING Unplug the monitor from the AC power source and detach all accessories before cleaning Do not immerse the unit in water or allow liquids to...

Страница 25: ...ng solution follow the manufacturer s directions carefully to avoid damaging the transducer 3 Wipe the transducer with a cloth soaked in water 4 Wipe the remained humidity with clean and dried cloth 6...

Страница 26: ...tor Note The manufacturer has no responsibility for the effectiveness of controlling infectious disease using these chemical agents Please contact infectious disease experts in your hospital for detai...

Страница 27: ...afety in Presence of Flammable Gases not suitable for use in presence of flammable gases Electromagnetic Compatibility Group Class A Mode continuous work Power Supply Working Voltage AC 100V 240V Freq...

Страница 28: ...on the operation of pregnant women there will be a mark display in the bottom area of FHR wave display section FHR Alarm Alarm for high and low FHR which exceeds appointed limit A 1 2 Transducers 1 Ul...

Страница 29: ...of FHR Symptom Possible Cause Solution Inconstant trace Inconstant display Wrong FHR No The pregnant woman is too fat No Improper ultrasound transducer position Change the position of ultrasound tran...

Страница 30: ...rately and neither too tight nor too loose Maternal Movement Relax patient s spirit Fetal Movement Wait for a moment then monitor Too high TOCO sensitivity higher than 100 unit The body pressure from...

Страница 31: ...Economic Technical Development Zone Hebei Province PRC Tel 86 335 8015433 Fax 86 335 8015432 E mail cms contecmed com cn Website http www contecmed com cn EC REPRESENTATIVE Shanghai International Hold...

Отзывы: