![Contec CMS800G Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/contec/cms800g/cms800g_user-manual_2655988029.webp)
2 Noises
Symptom
Possible Cause
Solution
Noise
Too high volume sets
Adjust the volume down
Interfered by handset or other
interfering source
Keep the handset or other
interfering source far away
3 Recorder Errors
Symptom
Possible Cause
Solution
Paper jam
Wrong feeding paper or paper is
affected with damp
Feed paper correctly and keep
paper from moist
Recorder does not work
button is disabled
Press the
button again
Out of paper
Feed paper
Just push print button, the printing of
last line not finished.
Waiting until it is finished.
4 Ultrasound Monitoring of FHR
Symptom
Possible Cause
Solution
Inconstant trace
Inconstant display
Wrong FHR
No
The pregnant woman is too fat
No
Improper ultrasound transducer
position
Change the position of
ultrasound transducer
Loose abdomen belt
Tighten abdomen belt
Superfluous coupling gel
Wipe off superfluous coupling
gel
Fetal movement
Wait for a moment then monitor
Maternal movement
Relax patient’s spirit
Inadequate coupling gel
Use recommended coupling gel
quantity
Doubtful FHR
Record Fetal heart rate wrongly
Change the position of
ultrasound transducer
The ultrasound transducer is not
placed well on the abdomen, and
the mixed noise has been
recorded
Change the position of
ultrasound transducer
Feint trace or no trace
Improper paper
Use the paper recommended
by manufacturer