Figure 3.11Print Setup
Revolving the knob key to setup print function. The Corresponding function and the adjustable
ranges are showed in the table3.5:
Table 3.5 The print setup function and adjustable ranges
No
Function
Adjustable Range
1
SPEED(Print
Speed)
1cm/min, 2cm/min, and 3cm/min. The default value is 3cm/min.
2
LENGTH(Print
Length)
0④24(hours) ,the interval is 10 minutes.
The default value is 30 minutes.
73
BASELINE(Baselin
e offset)
Adjustable:-10~+10, the interval is 1, the default value is0.
4
BASELINE PRINT Print the ladder-from testing wave.
5
PAPER(Selection of
printing paper type)
Please select it between American standard printing paper and
Europe standard printing paper.
6
EXIT
Return to the upper menu.
4 Installation
Note:
To
ensure
that
the
monitor
works
properly
,
please
read
this
chapter
and
Chapter 1 Safety
Guidance
. And follow the steps before using the monitor.
4.1 Open the Package and Check
Open the package and take out the monitor and accessories carefully. Put the monitor at safe and
reliable place. Check the components according to the packing list.
Check for any mechanical damage.
Check all the cable, and accessories.
If there is any problem, contact us for your local distributor immediately.
4.2 Connect the Power Cable
Make sure the AC power supply of the monitor complies with the following specification:
100V-240AV, 50/60Hz.
Consider the local power supply range, if the power supply of the monitor exceeds the range,
please add regulator equipment.
Apply the power socket of the monitor. Plug one end of the power cable to the power socket
of the monitor. Connect the other end of the power cable to a grounded 3-phase power output
special for hospital usage.
Connect the ground wire if necessary.