background image

 

CONTENTS 

Meter Design goals 

Meter Operation Basics 

Background Noise Tests 

Program RMS/Peak Levels 

Evaluating Frequency Response 

Headphone/Flat spectrum output 

Line In Flat Response 

Meter Specifications / Calibration 

 

Thank you for choosing our Field Strength Meter. It has been designed 

to ensure that Hearing Loop (Induction Loop) Systems be installed and certified 
to the new IEC 60118-4 standard. When these standards are met the user expe-
rience will be of equal quality wherever they have the opportunity to use their T-
coils. Our design team interviewed many installers throughout the industry which 
resulted in the following design goals: 

 

1. Very accurate and easy to read 

 

2. Simple/straightforward operation 

 

3. Based on the IEC 60118-4 specification 

 

4. Headphone output 

 

5. Flat spectrum output  

 

6. Built to last and be reliable 

As you follow through this manual, we will review how the meter should be used 
to certify that the requirements of the IEC specification have been met. 

Again, Thank you  

Contacta Design and Engineering Team 

Meter Design Goals 

 

Meter Specifications:

 

Measurement Range:   

 -62dB to +9dB (0dB = 400mA/m) 

True-RMS Crest Factor:  

<3  

Resolution:  

 

 

<0.1dB resolution for levels  

 

 

 

 

 

over –32dB 

Displayed resolution: 

 

0.1 dB  

Detection Type:  

 

True RMS on all features  

Sensor:  

 

 

Pickup Coil  

Direction of Sensitivity:    

Length-wise of meter’s longer dimension 

 

 

 

 

and Parallel to meter face (Position noted 

 

 

 

 

on Product)  

Calibration:  

 

 

Calibrated at 1,000 Hz (sine) to read 0  

 

 

 

 

dB at 5.03 mG  

Frequency Response:    

Flat ±1dB from 50 Hz to 10,000 Hz  

A-Weighted:  

 

 

Class 2 meter specified in IEC 61672-1  

Power Source:   

 

Single 9v Battery & External Power  

 

 

 

 

Jack  

Headphone Jack:  

 

Output A-Weighted or Flat selectable 

Display:  

 

 

16x2 LCD Character Display  

Backlight:  

 

 

Variable brightness blue LED  

 

 

Calibration and Warranty: 

This meter has been calibrated with equipment that is traceable back to the National Bureau 
of Standards. Under general use this meter should not require regular calibration, however if 
you return the unit to Contacta, the calibration can be checked and supplied with a calibration 
curve. Please call to get a service number and check the costs involved.  

Contacta, Inc.     332 E. Lakewood Blvd #400, Holland, MI 49424  Phone: (616) 392-3400 
Email: [email protected]

This meter is warranted to be free of component, design and manufacturing defects for a 
period of one year. Contacta accepts no responsibility for system designs or installations 
where this meter was used for test and calibration purposes.  Please contact us if you have 
any questions. Thank you. 

Содержание Loop-FSM

Страница 1: ...Model LOOP FSM User Manual Audio Frequency Induction Loop Field Strength Meter ...

Страница 2: ...ion 0 1dB resolution for levels over 32dB Displayed resolution 0 1 dB Detection Type True RMS on all features Sensor Pickup Coil Direction of Sensitivity Length wise of meter s longer dimension and Parallel to meter face Position noted on Product Calibration Calibrated at 1 000 Hz sine to read 0 dB at 5 03 mG Frequency Response Flat 1dB from 50 Hz to 10 000 Hz A Weighted Class 2 meter specified in...

Страница 3: ...p your own adapter use the following information 1 8 in TRS connector XLR female connector Tip Pin 2 Ring Pin 3 Sleeve Pin 1 3 Meter Operation Basics Turning the meter ON and Off On Press and hold button A until the screen lights Off Press and continue to hold button A until the Powering Down process is finished and the screen goes dark Selecting the meter mode When you press Button A it first dis...

Страница 4: ...k noise signal Cont Record the RMS level readings for each of the frequencies As in method I if the level does not vary by more than 3dB the sys tem as installed will meet the IEC specification This method was requested by the field engineer so they could run the test with one instrument without continually adjusting the frequency source It also makes it easy to conduct the test in more than one l...

Страница 5: ...o set the field strength to an average of 12dBA as it will constantly vary based on the program used to set up the system We will use the PEAK read ing to confirm that both our design and equipment meet the IEC specification Pressing button B resets the PEAK reading How to test then adjust the loop system level Once the hearing loop system has been installed play some bandwidth limited 100Hz to 5K...