Background Noise Level Test 

 
 

 

 

After turning the meter on it comes up in the Background 

Noise “A weighted” measurement mode. The display will indicate 
the RMS reading on the first line and the MAX reading on the second 
line. Pressing button “B” resets the MAX reading. 

 
 
 
 

How to test for background noise: 
 

When testing a new facility for background noise first turn on 

all lights, fans, sound system and other electrical equipment which 
is normally on when the facility is in use. Usually the loop system has 
not been installed yet, however if you are certifying an installation 
this test is done without the hearing loop system turned on. 
 

Walk throughout the seating area where the loop system will 

be used, holding the meter in a vertical position at the listening 
plane height. The important reading will be the MAX reading, Howev-
er it is important to watch the RMS reading. If the MAX reading ex-
ceeds –32dBA (readings above –32dBA will have a lower negative 
number ex. –30dBA means there is more background noise than       
–35dBA), you will need to document the areas where those higher 
noise levels are found.  

 

IEC 60118-4 Notes and Requirements 
 

The standard as revised in 2004 notes that any background 

noise level lower than –47dBA will result in a excellent  signal to 

noise ratio, however levels below –32dBA are acceptable and do 
meet the requirements of the standard. If the background noise lev-
el is above –32dBA, then the facility management should be notified 
and the source of the interference found and repaired. 

 

Method II  -  Pink noise signal Cont. 

 

 

 
 

 

Record the RMS level readings for each of the frequencies`. 

As in method I if the level does not vary by more than ±3dB the sys-
tem as installed will meet the IEC specification. This method was 

requested by the field engineer so they could run the test with one 
instrument without continually adjusting the frequency source. It 
also makes it easy to conduct the test in more than one location.

 

 

Headphone / Full Spectrum Output 

 

The headphone output jack serves two purposes: First it can 

be used to monitor the loop program and gives you an “A Weighted” 
output signal that can be listened to with  standard 1/8” stereo head-

phones. To change the headphone volume slide the Use/Menu 
switch to Menu and use the Mode button to advance to Headphone 
Volume. Pressing the Select button advances to the volume adjust 
screen where the top button raises the volume and the bottom but-
ton lowers it. Once adjusted sliding the switch back to Use will save 
the setting. 
 
 

A full (flat) spectrum output can be sent from this same con-

nector which could then feed a spectrum analyzer. This would show 
the signal level at the various frequencies and help to both confirm 
proper operation and asses the frequency and level of any interfer-
ences.  
 

To switch from  “A-Weighted” to Flat: With the meter turned 

on slide the Use/Menu switch over to Menu. Press the Mode button 
“A” once and you will advance to the Headphone Jack setup screen. 
Then pressing the Select button “B” you can choose either A-
Weighted or Flat Spectrum.  

 

Menu Adjustments 

In the Menu mode, selected by the slide switch, the following items 
can be adjusted: Backlight level, Headphone output type, Head-
phone volume, Display units (dB, mG, uT) and Power timeout (5 to 
30, or none. 

Содержание Loop-FSM

Страница 1: ...Model LOOP FSM User Manual Audio Frequency Induction Loop Field Strength Meter ...

Страница 2: ...ion 0 1dB resolution for levels over 32dB Displayed resolution 0 1 dB Detection Type True RMS on all features Sensor Pickup Coil Direction of Sensitivity Length wise of meter s longer dimension and Parallel to meter face Position noted on Product Calibration Calibrated at 1 000 Hz sine to read 0 dB at 5 03 mG Frequency Response Flat 1dB from 50 Hz to 10 000 Hz A Weighted Class 2 meter specified in...

Страница 3: ...p your own adapter use the following information 1 8 in TRS connector XLR female connector Tip Pin 2 Ring Pin 3 Sleeve Pin 1 3 Meter Operation Basics Turning the meter ON and Off On Press and hold button A until the screen lights Off Press and continue to hold button A until the Powering Down process is finished and the screen goes dark Selecting the meter mode When you press Button A it first dis...

Страница 4: ...k noise signal Cont Record the RMS level readings for each of the frequencies As in method I if the level does not vary by more than 3dB the sys tem as installed will meet the IEC specification This method was requested by the field engineer so they could run the test with one instrument without continually adjusting the frequency source It also makes it easy to conduct the test in more than one l...

Страница 5: ...o set the field strength to an average of 12dBA as it will constantly vary based on the program used to set up the system We will use the PEAK read ing to confirm that both our design and equipment meet the IEC specification Pressing button B resets the PEAK reading How to test then adjust the loop system level Once the hearing loop system has been installed play some bandwidth limited 100Hz to 5K...

Отзывы: