10.
Enter the desired power limit
and press
:
METHOD 1|1
POWER
150
W
This limit is automatically set to
maximum and cannot be changed in case
of a voltage gradient step:
11.
Enter the desired timer unit (h or kVh)
and press
:
METHOD 1|1
TIMER UNIT
h
12.
Enter the desired timer value
and press
:
METHOD 1|1
TIMER
00:30
h
The value format is 'hh:mm' when
entering a value in hours, a value with 2
decimals when entering kiloVolthours.
13.
Decide if the experiment should
automatically proceed with a next step,
press
:
METHOD 1|1
END METHOD
NO
Choose NO to resume at point
6
with the
next step,
choose YES to stop after this
step.
14.
Press
/
to start the experiment.
•
At any time, you can return to the main menu by holding
pressed.
•
At any time, you can avoid programming unnecessary parameters. To do so, simply press
/
instead of
after entering the last relevant value and the instrument will immediately start the experiment keeping the
other parameters to their previous values.
•
To ensure that the desired parameter is kept constant, we suggest to pre-set all other parameters to its
maximum.
•
A blinking unit of the controlling parameter appears when the regulation is complete.
•
The timer will always count down except when no timer value has been entered.
Rev. 1605