Manual Power Supplies

Frequently Asked Questions

What are the relations between Voltage, Current, Power and Resistance?

Power (W) = Voltage (V) x Current (A)
Resistance (

W

) = Voltage (V) / Current (A)

How does a power supply react after pressing RUN?

The internal generator will start building up the high voltage at the output terminals while voltage and 
current are constantly measured and power calculated. When one of the pre-set parameters is 
exceeded, the generator stops and will keep that parameter constant. 

How important is the resistance of an electrophoresis unit?

The resistance of an electrophoresis unit depends on its size, gel thickness, amount of buffer, buffer 
conductivity and temperature. This resistance will normally decrease in time due to a slowly increasing 

temperature. Electrophoresis units which have a resistance below the minimum load resistance of a 
power supply will trigger an alarm! Read the output voltage and current during a run to measure the 
resistance and use above formula to calculate the value.

How to keep a constant voltage during a run?

Program the desired voltage and a higher current and power then the maximum expected values:
Current > Voltage / Resistance
Power > Voltage x Current

How to keep a constant current during a run?

Program the desired current and a higher voltage and power then the maximum expected values:
Voltage > Current x Resistance
Power > Voltage x Current

How to keep a constant power during a run?

Program the desired power and a higher voltage and current then the maximum expected values:
Voltage > Current x Resistance
Current > Voltage / Resistance

Why are my output values different from those of a similar experiment?

Either your programmed parameters are not equal to those described or the resistance of your 
electrophoresis unit is different (see above). It cannot be due to e.g. an other model of power supply as 
the relations between Voltage, Current, Power and Resistance are monitored in the same way by any 
instrument (the electrical laws cannot be disregarded!).

What about connecting more than one unit to the same power supply?

The outlets being in parallel each electrophoresis unit will be supplied with exactly the same voltage. 

However, current and power may differ due to differences between them even when exactly the same 
model, gel, buffers, etc... are used. Therefore, it is recommended to run several electrophoresis units 

only in the constant voltage mode on the same power supply. 
Mind also that adding tanks will increase the current and the power. These parameters may be limited 

by the power supply and cause it to switch over to constant current or constant power.

                                                                                  

 

                                                                              

Rev. 1605

11

Содержание EV1450

Страница 1: ...MANUAL HANDLEIDING MODE D EMPLOI ANLEITUNG EV1450 EV2150 EV2310 EV2320 EV2650 EV3020 EV3330 EV3610 EV3620 February 2015...

Страница 2: ...trast 3 Date Time 4 Password 4 Detection 5 Data 5 Info 5 Method programming 6 Manual programming 8 Running condition 9 Display 9 Buttons 9 Adjusting parameters 9 End of procedure 10 Alarm messages 10...

Страница 3: ...sed under normal laboratory conditions and in accordance with the operating limitations and maintenance procedures in this instruction manual and when not having been subject to accident alteration mi...

Страница 4: ...safety cover 6 Never make any other connections such as e g putting several power supplies in series or in parallel 7 In order to prevent electric shock never open the back plate nor remove the cover...

Страница 5: ...preset the system settings of the power supply for language display contrast real time clock password detections data recording System Language The power supply has several languages built in for the...

Страница 6: ...TIME MODIFY 07 12 2015 13 08 47 System Password When desired a password can be set which will be requested when trying to modify the setup or the parameters of a procedure The user can then only run...

Страница 7: ...for the remaining time 1 Follow these steps to select LOW CURRENT and press PROCEDURE MANUAL METHOD SETUP SYSTEM DETECTION DATA INFO DETECTION LOW CURRENT POWER FAIL 2 Decide whether or not the output...

Страница 8: ...ctrophoresis tank to the instrument 3 Switch the instrument on 4 Select METHOD and press PROCEDURE MANUAL METHOD SETUP 5 Select the desired method and press CHOOSE METHOD 1 6 When previously multiple...

Страница 9: ...1 END METHOD NO Choose NO to resume at point 6 with the next step choose YES to stop after this step 14 Press to start the experiment At any time you can return to the main menu by holding pressed At...

Страница 10: ...lue and press MANUAL TIMER 00 30h The value format is hh mm when entering a value in hours The format is a value with 2 decimals when entering kiloVolthours Set to zero when required to continue until...

Страница 11: ...also used for system messages when required The unit of the parameter that is being kept constant is given with blinking bold characters Buttons The keys be used as follows Buttons for selecting anot...

Страница 12: ...e starting the experiment again GROUND LEAKAGE A dangerous ground leakage is present check thoroughly the complete assembly Your power supply is NOT defective See Frequently Asked Questions at the end...

Страница 13: ...ep a constant current during a run Program the desired current and a higher voltage and power then the maximum expected values Voltage Current x Resistance Power Voltage x Current How to keep a consta...

Страница 14: ...Manual Power Supplies Specifications Rev 1605 12...

Страница 15: ...fective equipment returned to it during the warranty period without charge CIF Turnhout prepaid by sender provided the equipment has been used under normal laboratory conditions and in accordance with...

Страница 16: ...EMC DIRECTIVE 2004 108 EC EN61326 1 ROHS DIRECTIVE 2011 65 EU EN50581 2012 Consort bvba Parklaan 36 B 2300 Turnhout Belgium Tel 32 0 14 41 12 79 Fax 32 0 14 42 91 79 Name Jan De Ceuster Signature Tit...

Отзывы: