Manual Power Supplies
Manual programming
Manual programming permits to set voltage, current, power and time limits for a simple routine
electrophoresis run. Parameters can be changed temporarily without interrupting the run.
1.
Verify if the instrument is switched off.
2.
Connect the electrophoresis tank to the instrument.
3.
Switch the instrument on.
4.
Select
MANUAL
and press
:
PROCEDURE?
MANUAL
METHOD
SETUP
5.
Enter the desired voltage limit and press
:
MANUAL
VOLTAGE
300
V
6.
Enter the desired current
limit
and press
:
MANUAL
CURRENT
500
mA
7.
Enter the desired power
limit
and press
:
MANUAL
POWER
150
w
8.
Enter the desired timer unit (h or kVh)
and press
:
MANUAL
TIMER UNIT
h
9.
Enter the desired timer value
and press
:
MANUAL
TIMER
00:30
h
The value format is 'hh:mm' when entering a
value in hours. The format is a value with 2
decimals when entering kiloVolthours. Set to
zero when required to continue until
/
is
pressed or the power is switched off.
10.
Press
/
to start the experiment. See
Running Condition
.
•
At any time, you can return to the main menu by holding
pressed.
•
At any time, you can avoid programming unnecessary parameters. To do so, simply press
/
instead of
after entering the last relevant value and the instrument will immediately start the experiment keeping the
other parameters to their previous values.
•
To ensure that the desired parameter is kept constant, we suggest to pre-set all other parameters to its
maximum.
•
A blinking unit of the controlling parameter appears when the regulation is complete.
•
The timer will always count down except when no timer value has been entered.