E-45
E
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
d
= [3.5 / E1]
√
P
80 MHz to 800 MHz
d
= [7 / E1]
√
P
800 MHz to 2.5 GHz
Where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
,
should be less than the compliance
level in each frequency range
b
.
Interference may occur in the
vicinity of equipment marked with
the following symbol.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and/or people.
Table 4: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The
ConMed Linvatec
VP47XX
and
VP47XXF
Monitors are intended for use in the
electromagnetic environment specified below. The customer or the user of this
VP47XX
/
VP47XXF
Monitor should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environment
guidance
Содержание Linvatec VP4726
Страница 54: ...E 50 E ...
Страница 58: ...F iv Table des matières Page F ...
Страница 114: ...F 56 F ...
Страница 118: ...G iv Inhaltsverzeichnis Seite G ...
Страница 170: ...G 52 G ...
Страница 174: ...I iv Sommario Pagina I ...
Страница 228: ...I 54 I ...
Страница 232: ...S iv Contenido Página S ...
Страница 285: ......