Recommended Annual Monitor Maintenance Requirements
Table 1: Recommended Annual Monitor Maintenance Requirements
Test Type
Test Value
Ground Impedance
ZG < 100 mOhm from the ground pin on the power inlet module
to the Console’s exposed metal parts.
*
Test Chassis Leakage Currents
IL < 100 uA in NORMAL Condition.
IL < 500 uA in Single Fault Conditions [300 uA US deviation]
*
Test Earth Leakage Currents
IL < 500 uA NORMAL Condition [300 uA US deviation]
IL < 1 mA Single Fault Condition
*
Test Dielectric Withstand
Test Line and Neutral to Ground @ V = 1500 V~, no breakdown
*
* See IEC 60601-1 for test methods.