background image

24k

© 2010 Linvatec Corporation, a subsidiary of C

on

M

ed

 Corporation, 01/2010, COS 6024 Rev. 2

Optional Operational Features

The Clearfield 

button combines 

tamponade and 

lavage modes 

for a safe, 

fast response 

to blood and 

debris in the 

operative field.

»  Tap to raise pressure in the joint by 

20mmHg to control bleeders.

»  Hold to increase the flow rate to 

quickly clear the joint of debris and 

blood to maintain visibility.

The Drain 

function 

allows 

surgeons 

to control 

the rapid 

removal of 

fluid form the 

joint.

»  Press and hold to activate drain.
»  Drain function continues until 

button is released.

The Declog 

function 

reverses 

suction to 

remove debris 

from the shaver 

blade.

» Remove shaver from the joint.
» Turn shaver on.
» Press shaver Declog button.

Controlling Outflow and Suction Rates

Press the Outflow and Shaver Suction buttons on the front panel or remote control to toggle the rates between off, low, 

medium, and high.

Outflow 

Shaver Suction

Low = 50 ml/min 

Low = 500 ml/min

Medium = 150 ml/min 

Medium = 850 ml/min

High = 300 ml/min 

High = 1200 ml/min