UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
2.0
Operation
2.1
Unpacking and Inspection
When received, carefully remove the instrument from its shipping container. A visual inspection of the
instrument’s external surfaces should be performed immediately after unpacking. If obvious damage has been
incurred during transit, the shipping agency and the distributor should be notified as soon as possible.
Instructions as to how to proceed after assessment of the damage will then be determined. If the instrument
shows no signs of damage, check to be sure all the required equipment and accessories have been received. See
“Options, Replacement Kits” on page 54, for all options.
NOTE:
Prior to operating battery models, reinstall (+) cable, red wire, on battery.
2.2
Standard Display/Keyboard Front Panel Functions
UPS3000 See Figure 2-1 on page 5, UPS3110 See Figure 2-2 on page 5, UPS3210 See Figure 2-3 on page 6.
1. Six Digit LED display of the measured pressure. This display also prompts if a problem occurred in
performing an operation.
2. Three Annuciators (4). Two of the LEDs are used for indicating whether the instrument is operating as a
Gage pressure instrument or as an absolute (ABS) pressure instrument. The third LED indicates the
conversion units that the pressure data is currently being displayed in.
3.
ZERO
function key (1). In
GAGE
mode and ABS only instruments this key allows the indicator to be
zeroed.
NOTE:
If attempting to perform a push to ZERO while the pressure data is in motion, the instrument will not go to ZERO
.
NOTE:
The four, or five if options have been installed, push button switches located across the lower face of the front
panel are individually actuated by momentarily pressing with a light fingertip motion anywhere within the black outlines
of the button. When actuated, each switch produces a visual feedback through the digital display.
4.
ABS/GAGE
function key (2), toggles instrument to be a Gage pressure instrument or an absolute (ABS)
pressure instrument.
NOTE:
This key is blank and non-functional if unit is an Absolute (Abs) only pressure unit or a Gage only pressure unit
.
5.
CONV
conversion key (3), is used to select the conversion units that the pressure data is currently being
displayed in.
6.
CAL
function key (5), activates the internal Self-Check feature. See “GAGE Mode Self-Check” on
NOTE:
CAL key is non-functional in absolute mode
.
7.
RANGE SELECT
switch (6), used to select the desired pressure range.
NOTE:
Do not turn switch between ranges during a cycle.
8. UPS3110 instrument with maximum range of 2K PSI or less:
PRESSURE LIMIT CONTROL
(7), for
regulating the input pressure to the unit.
9. UPS3110 instrument with maximum range of 2K PSI or less:
PRESSURE LIMIT MONITOR
(8), indicates
input pressure going to the unit.
10. UPS3110 instrument:
TEST PORT
(9), for connecting to the Device-Under-Test (DUT).
11. UPS3110 instrument:
INPUT
Valve (10), used for nearing target pressure of the DUT.
12. UPS3110 instrument:
VERNIER
Valve (11), used for precisely setting target pressure of the DUT.
13. UPS3110 instrument:
VENT
Valve (12), used for venting pressure of the DUT.
Содержание UPS3000
Страница 2: ...2 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Страница 50: ...46 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Страница 53: ...48 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......