24
UPS3000/UPS3110/UPS3210 Operation and Maintenance Manual
3.7.1
For UPS3000/UPS3210 Absolute/Gage Switch Selectable Units ONLY
If the current barometric pressure is
below
14.7 PSIA, the offset is positive, see Example 1. If the current
barometric pressure is above 14.7 PSIA, the offset is negative, see Example 2. If the current barometric pressure
is 14.7 PSIA, then no offset is needed.
Example 1:
If the current barometric pressure is
lower
than 14.70 PSIA, subtract the current barometric pressure from 14.70.
14.70 PSI: UPS3110 reference point
-14.55 PSI: Current barometric pressure
0.15 PSI: Positive Delta Offset
Complete the following steps
NOTE:
UPS3000 refer to Figure 2-1 on page 5 and Figure 2-4 on page 6. UPS3210 refer to Figure 2-3 on page 6 and
1. Connect Test Standard to UPS3000/UPS3210 Input Port similar to Figure 3-1 on page 19.
2. Open the vent valve connected between Test Standard and UPS3000/UPS3210.
3. Using the
RANGE SELECT
switch (6), select the lowest pressure range on the UPS3000/UPS3210.
4. Press the
ENTER
push-button, as shown in Figure 3-2 on page 20, on the UPS3000/UPS3210. The
display reads zero. Repeat this step for the middle and high ranges.
5. Close the vent
valve connected between Test Standard and UPS3000/UPS3210. Select the lowest
pressure range of the UPS3000/UPS3210.
6. Input a pressure into the UPS3000/UPS3210 until the display reads the value of the Positive Delta
Pressure, 0.15 PSI as in this example.
7. Depress the
ENTER
push-button on the UPS3000/UPS3210. The display reads zero. Without touching the
input pressure
repeat this step for the middle and high ranges.
8. Using the
STEPPER
push-button, as shown in Figure 3-2 on page 20, place the unit into
SHUNT RESISTOR
CALIBRATION
mode. The display will be as shown in Figure 3-5 on page 23.
9. With UPS3000/UPS3210
RANGE SELECT
switch (6) switched in the high range, depress the
ZERO
push-button on the front panel. Verify that the display reads 0.00. Upon release of the push-button the
display will be as shown in Figure 3-5 on page 23.
10. Depress the
ENTER
push-button. The display will respond with
“[] [] [] [] [] []”
until the button is
released.
11. Pressing the
STEPPER
push-button bring the indicator back to its initial Data Recall display condition as
shown in Figure 3-2 on page 20. The display shows
1 XXX
(three arbitrary digits).
12. Depress the
DATA ENTER
switch, S2, located approximately in the middle of the circuit board. See Figure
2-7 for the switch location.
13. If the data is accepted, the three-digit number on the right side of the display will indicate
377
for as long
as the
DATA ENTER
switch, S2, is depressed.
14. The barometric offset is now complete and
CALIBRATE/OPERATE
switch, S3, must be returned to its
normal operating positions as shown in Table 3-1 on page 20. The Test Standard may now be
disconnected.
Содержание UPS3000
Страница 2: ...2 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Страница 50: ...46 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Страница 53: ...48 UPS3000 UPS3110 UPS3210 Operation and Maintenance Manual...
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......