Guidance and manufacturer’s declaration – electromagnetic immunity

Emission test     IEC 60601 test level    Compliance level    Electromagnetic environment – guidance

The device is intended for use in the electromagnetic environment specified below. The user of the device 

should assure that it is used in such an environment.

NOTE   U

T

 is the a.c. mains voltage prior to application of the test level.

Electrostatic 

discharge (ESD)

IEC 61000-4-2

Electrical fast 

transient/burst

IEC 61000-4-4

Surge

IEC 61000-4-5

Voltage 

dips, short 

interruptions 

and voltage 

variations on 

power supply 

input lines

IEC 61000-4-11

Power 

frequency 

(50Hz/60Hz) 

magnetic field 

IEC 61000-4-8

Floors should be wood, concrete or 

ceramic tile. If floors are covered with 

synthetic material, the relative humidity 

should be at least 30%.
Mains power quality should be that 

of a typical commercial or hospital 

environment.

Mains power quality should be that 

of a typical commercial or hospital 

environment.

Mains power quality should be that of a 

typical commercial or hospital environment. 

If the user of the device requires continued 

operation during power mains interruptions, 

it is recommended that the device be 

powered from an uninterruptible power 

supply or a battery.

Power frequency magnetic fields should 

be at levels characteristic of a typical 

location in a typical commercial or hospital 

environment.

±6 kV contact

±8 kV air

±2 kV for power 

supply lines

±1 kV for input/

output  lines
± 1 kV line(s) to 

line(s)

± 2 kV line(s) to earth

3 A/m

3 A/m

±6 kV contact

±8 kV air

±2kV for power 

supply lines

±1 kV differential 

mode

<5% U

T

 

(>95% dip in U

T

)

for 0.5 cycle
40% U

T

(60% dip in U

T

)

for 5 cycles
70% U

T

(30% dip in U

T

)

for 25 cycles
<5% U

T

 

(>95% dip in U

T

)

for 5 sec

<5% U

T

 

(>95% dip in U

T

)

for 0.5 cycle
40% U

T

(60% dip in U

T

)

for 5 cycles
70% U

T

(30% dip in U

T

)

for 25 cycles
<5% U

T

 

(>95% dip in U

T

)

for 5 sec

Содержание Theralief TN100

Страница 1: ...TENS ELECTROTHERAPY FOR TARGETED PAIN RELIEF INSTRUCTION MANUAL All TN100 Models ...

Страница 2: ......

Страница 3: ...rience temporary relief from aches and pains in different parts of their bodies INDICATIONS FOR USE To be used for temporary relief of pain associated with sore and aching muscles in the shoulders waist back arms and legs due to strain from exercise or normal household and work activities SAFETY INFORMATION AND WARNINGS CONTRAINDICATIONS Do not use this device on users who have a cardiac pacemaker...

Страница 4: ... following conditions Across the chest because the introduction of electrical current into the chest may cause rhythm disturbances to the heart which could be lethal Over painful areas Please consult with your physician before using this device if you have painful areas Over open wounds or rashes or over swollen red infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis varico...

Страница 5: ...mmediately and report to the distributor seller 11 The long term effects of electrical stimulation are unknown 12 Since the effects of stimulation of the brain are unknown stimulation should not be applied across the head and electrodes should not be placed on opposite sides of the head 13 The safety of electrical stimulation during pregnancy has not been established 14 Some users may experience s...

Страница 6: ...Fragile handle with care Keep dry Dispose of the battery containing device according to local state or federal laws Properly recycle Normal working ambient temperature 5 C 40 C 23 F 104 F Normal working ambient humidity 80 Store and transport ambient temperature 10 C 50 C 14 F 22 F Store and transport ambient humidity 90 CONTROL UNIT GEL ELECTRODE PAD POWER ...

Страница 7: ...d Then stick the black side of the pad to your skin in an area where you feel pain Please re read the safety information and warnings for conditions under which you should not use the TENS unit 4 Press the ON OFF button to turn on the power 5 Select one of the stimulation modes 6 Adjust the stimulation intensity 7 Enjoy the treatment 8 The unit will turn off after 30 minutes of use To manually tur...

Страница 8: ...s STEP 3 PLACE THE ELECTRODE PAD WITH THE CONTROL UNIT SNAPPED IN PLACE ONTO THE TREATMENT AREA SUCH AS YOUR ARM YOUR BACK ETC You can start by applying the pad directly over an area where you experience pain However you may also want to experiment by moving the pad to adjacent areas Pain in one area may be caused by a tight muscle in another area for instance After using the device for a while yo...

Страница 9: ...level that feels comfortable to you NOTE With the increase of intensity total of 20 levels you may experience sensations such as tingling or vibration Therefore gradually increase the intensity and stop increasing when a comfortable level is reached You may prefer different levels of intensity in different modes on different days and on different parts of the body STEP 7 STIMULATION DURATION After...

Страница 10: ... against your skin in the treatment area and secure the wrap with the elastics wrapped around your body part for instance your knee Then adhere the part of the elastics against the fuzzy side of the wrap 4 Follow directions STEP 4 etc above REMOVAL AND STORAGE 1 Turn the device off before removing the electrode pad from your skin 2 Lift at the edge of the electrode pad and slowly peel 3 Replace cl...

Страница 11: ...ctive Contact Conair for replacements at 1 800 3 CONAIR Do not let the sticky side of the pad touch anything other than your skin Be careful not to handle with greasy fingertips Power Source 3 7V Battery Number of Output Modes 5 Timer Range minutes 30 Dimensions mm L x W x D 188x85x10 Waveform Biphasic Shape Rectangular Maximum Output Voltage 66V 500Ω Maximum Output Current 132mA 500Ω Pulse Durati...

Страница 12: ...er a replacement You can order a replacement at www conair com or call Conair for replacements at 1 800 3 CONAIR SKIN TURNS RED Stop treatment and move pads to another area If problem persists stop using the device and contact your physician ELECTROMAGNETIC COMPATIBILITY 1 This product needs special precautions regarding electromagnetic compatibility EMC and needs to be installed and put into serv...

Страница 13: ... fluctuations flicker emissions IEC 61000 3 3 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The device is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buil...

Страница 14: ... should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of the device requires continued operation during power mains interruptions it is recommended that the device be powered from an uninterruptible power supply ...

Страница 15: ...nnot be predicted theoretically with accuracy continued next page Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 6 Portable and mobile RF communications equipment should be used no closer to any part of the device including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Where P is the m...

Страница 16: ...s and people 0 01 0 1 1 10 100 To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnormal performance is observed additional measures may be ne...

Страница 17: ...ation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit ...

Страница 18: ...e of manufacture ANY IMPLIED WARRANTIES OBLIGATIONS OR LIABILITIES INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE SHALL BE LIMITED IN DURATION TO THE 12 MONTH DURATION OF THIS WRITTEN LIMITED WARRANTY Some states do not allow limitations on how long an implied warranty lasts so the above limitations may not apply to you IN NO EVENT SHALL C...

Страница 19: ......

Страница 20: ...roduct at www conair com registration SERVICE CENTER CONAIR CORPORATION Service Department 7475 North Glen Harbor Blvd Glendale AZ 85307 Visit us on the web at www conair com 2016 Conair Corporation stock image Model TN100 15PS018177 ...

Отзывы: