8. Technical data 
 

Type: 5306 

500 

Manufacturer: 

Compur Monitors, München 

Protection class: 

EEx ib IIC T6 

Operating environment:  

II 2 G 

Operating temperature:  

-20°C to +50°C ( -4°F to + 122°F) 

Power supply: 

 

2 x Li, 3V:  

 

 

only CR 2477N (Renata) or CR 2477 (Panasonic) 

Sensor principle 

Electrochemical 

Humidity   

typ.: 20% to 90% r.h. 

Pressure 

920 hPa – 1120 hPa 

Display 4-digit 

LCD-Display 

Weight 

130 g  (4,6 oz) 

Dimensions (With filter cap) 

4,9 x 2,6 x 1“ 

Battery lifetime  

about 800 hours 

Alarm thresholds adjustable range  0 – 100% of range 
Alarms loudness 

typ.: 80 db (A) / 30 cm 

Terminal Earphone 
Sensor warranty 

6 to 12 months 

 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
Specifications are subject to change without notice, and are provided only for comparison of products. 
The conditions, under which our products are used, are beyond our control. Therefore, the user must 
fully test our products and/or information to determine suitability for any intended use, application, 
condition or situation. All information is given without warranty or guarantee. Compur Monitors disclaims 
any liability, negligence or otherwise, incurred in connection with the use of the products and information. 
Any statement or recommendation not contained herein is unauthorized and shall not bind Compur 
Monitors. Nothing herein shall be construed as a recommendation to use any product in conflict with 
patents covering any material or device or its use. No licence is implied or in fact granted under the 
claims of any patent. Instruments are manufactured by Compur Monitors GmbH & Co. KG, Munich. 
The General Conditions of Supply and Service of Compur Monitors GmbH & Co. KG are applicable.

 

 

 
 
 
 
 
 
 
 

  5306 500 998 07 00 / 05.05                                                                                                                                          506110 

 

 

 

Содержание 5306 500

Страница 1: ...Operating Instructions Compur Monitox plus Subject to changes without notice Copyright Compur Monitors M nchen Editor Compur Monitors GmbH Co KG Weissenseestr 101 D 81539 M nchen...

Страница 2: ...of device certified for group II category 2 Its intended use is the measurement of toxic gases or oxygen Designed with increased safety it is applicable in zone 1 and zone 2 Please observe the followi...

Страница 3: ...ntly high concentrations z Wear the instrument near the breathing zone using the clip or neck chain z The filter cap should be protected from dripping water and exposure to excessive dust z The Monito...

Страница 4: ...ve 2 1 Gas calibration Place Monitox plus gas adapter part 569747 securely on the filter cap Connect span gas NOTE If using HCN Cl2 or other corrosive gases use non porus tubing such as Tygon lined wi...

Страница 5: ...20 9 oxygen Zero can be adjusted with pure Nitrogen GCAL or electronically ICAL 3 Alarm thresholds Compur Monitors recommends to strictly observe the local regulations for the alarm thresholds Open th...

Страница 6: ...t calibration 4 3 Battery replacement Open the instrument Disconnect and remove the battery pack Remove battery pack cover Replace the batteries Observe the proper polarity as is illustrated on the ba...

Страница 7: ...ion data Temperature is out of range This message is displayed for a short time interval Then the Monitox will go on to operate z The instrument operates out of the temperature compensated range Bring...

Страница 8: ...libration Current calibration Intermediate display when the operation mode has been altered and during the self test Self test in process Battery pre alarm Batt and the actual measured value alternate...

Страница 9: ...nder which our products are used are beyond our control Therefore the user must fully test our products and or information to determine suitability for any intended use application condition or situat...

Страница 10: ...10...

Отзывы: