4. Maintenance
4.1 Filter cap replacement
Carefully turn the old filter cap 90° counter clockwise. Insert the new filter cap. Turn 90°
clockwise. Make sure to use only filter caps for the relevant gas!
4.2 Sensor replacement
Remove filter cap. Pull sensor out. Some sensors are shipped with a plug that short circuits the
contacts. Remove this. Check best before date of the sensor. Note the output current, which is
on the sensor label. Perform a current calibration.
4.3 Battery replacement
Open the instrument. Disconnect and remove the battery pack. Remove battery pack cover.
Replace the batteries. Observe the proper polarity, as is illustrated on the battery pack cover.
Close, connect and mount the battery pack. Perform a functional test.