4. Maintenance 
 
4.1 Filter cap replacement 

Carefully turn the old filter cap 90° counter clockwise. Insert the new filter cap. Turn 90° 
clockwise. Make sure to use only filter caps for the relevant gas! 
 
 

4.2 Sensor replacement 

Remove filter cap. Pull sensor out. Some sensors are shipped with a plug that short circuits the 
contacts. Remove this. Check best before date of the sensor. Note the output current, which is 
on the sensor label. Perform a current calibration. 
 
 

 

 

 

 

 

 
 
4.3 Battery replacement 

Open the instrument. Disconnect and remove the battery pack. Remove battery pack cover. 
Replace the batteries. Observe the proper polarity, as is illustrated on the battery pack cover. 
Close, connect and mount the battery pack. Perform a functional test. 
 
 
 
 
 
 

5. Accessories  

 

 

 

 

 

 

 

 

 

 

 

Article No.: 

Earphone 

         566032 

Leather 

Pouch 

        503746 

Steel clip  

 

 

 

 

 

 

 

 

           568434 

Gas 

adapter 

         569747 

Suspender Clip 

 

 

 

 

 

                                554566 

Belt Clip   

 

 

 

 

 

 

                                568434 

Gas generator H

2

S   

 

 

 

 

 

 

 

510329 

Gas generator COCI

2

 

 

 

 

 

 

 

 

510634 

Gas generator HCN 100 ppm 

 

 

 

 

 

 

510717 

Gas generator HCN 20 ppm 

 

 

 

 

 

 

510640 

Gas generator NO

2

  

 

 

 

 

 

 

 

510741 

Gas 

generator 

CO 

        510790 

Gas generator CI

2

   

 

 

 

 

 

 

 

510204 

Gas generator SO

2

  

 

 

 

 

 

 

 

510279  

 
 

 

 

Содержание 5306 500

Страница 1: ...Operating Instructions Compur Monitox plus Subject to changes without notice Copyright Compur Monitors M nchen Editor Compur Monitors GmbH Co KG Weissenseestr 101 D 81539 M nchen...

Страница 2: ...of device certified for group II category 2 Its intended use is the measurement of toxic gases or oxygen Designed with increased safety it is applicable in zone 1 and zone 2 Please observe the followi...

Страница 3: ...ntly high concentrations z Wear the instrument near the breathing zone using the clip or neck chain z The filter cap should be protected from dripping water and exposure to excessive dust z The Monito...

Страница 4: ...ve 2 1 Gas calibration Place Monitox plus gas adapter part 569747 securely on the filter cap Connect span gas NOTE If using HCN Cl2 or other corrosive gases use non porus tubing such as Tygon lined wi...

Страница 5: ...20 9 oxygen Zero can be adjusted with pure Nitrogen GCAL or electronically ICAL 3 Alarm thresholds Compur Monitors recommends to strictly observe the local regulations for the alarm thresholds Open th...

Страница 6: ...t calibration 4 3 Battery replacement Open the instrument Disconnect and remove the battery pack Remove battery pack cover Replace the batteries Observe the proper polarity as is illustrated on the ba...

Страница 7: ...ion data Temperature is out of range This message is displayed for a short time interval Then the Monitox will go on to operate z The instrument operates out of the temperature compensated range Bring...

Страница 8: ...libration Current calibration Intermediate display when the operation mode has been altered and during the self test Self test in process Battery pre alarm Batt and the actual measured value alternate...

Страница 9: ...nder which our products are used are beyond our control Therefore the user must fully test our products and or information to determine suitability for any intended use application condition or situat...

Страница 10: ...10...

Отзывы: