6. Error messages
Whenever a malfunction occurs, the Compur Monitox plus will give you an error message.
The sensor current obtained during gas calibration is out of range.
z
Calibration gas concentration must be within 0,5 TLV and full scale.
z
Switch instrument off and on again to make it operate with the old calibration
data.
z
Sensor defective: Replace sensor.
The switch has been moved into the “ON”- position during calibration. The
instrument stops working.
z
Switch instrument off and on again to make it operate with the old calibration
data.
Temperature is out of range. This message is displayed for a short time
interval. Then the Monitox will go on to operate.
z
The instrument operates out of the temperature - compensated range. Bring it
into the specified temperature.
EEPROM defective. The instrument stops operating.
z
The instrument must be serviced.
The A/D-Converter input voltage is out of range. The instrument stops
working.
z
Switch instrument off and on again.
z
If error persists, check sensor.
z
If the sensor is OK and the problem persists, the instrument must be serviced.
Sensor is missing, has bad contact or is defective.
z
Replace sensor.
z
Check contacts.
z
If no success, replace sensor.
The Monitox
NO
2
0-10 ppm, NO
2
0-50 ppm and CI
2
do not have this feature.