6. Error messages 

Whenever a malfunction occurs, the Compur Monitox plus will give you an error message. 
 
 

 

The sensor current obtained during gas calibration is out of range. 

 

 

z

 Calibration gas concentration must be within 0,5 TLV and full scale.  

 

 

z

 Switch instrument off and on again to make it operate with the old calibration   

                         data. 

z

 Sensor defective: Replace sensor. 

 

 

 

                     

The switch has been moved into the “ON”- position during calibration. The 
instrument stops working. 

 

 

z

 Switch instrument off and on again to make it operate with the old calibration   

                         data. 
 

 

  

                     

Temperature is out of range. This message is displayed for a short time 
interval. Then the Monitox will go on to operate. 

 

 

z

 The instrument operates out of the temperature - compensated range. Bring it  

                         into the specified temperature. 
 

 

 

                      

EEPROM defective. The instrument stops operating. 

 

 

 

z

 The instrument must be serviced. 

 

 

 

                     

The A/D-Converter input voltage is out of range. The instrument stops 
working.

 

 

 

  

 

z

 Switch instrument off and on again. 

 

 

z

 If error persists, check sensor. 

 

 

z

 If the sensor is OK and the problem persists, the instrument must be serviced. 

 

 

 

                     

Sensor is missing, has bad contact or is defective. 

 

 

 

z

 Replace sensor. 

 

 

z

 Check contacts. 

 

 

z

 If no success, replace sensor. 

 

 

    The Monitox 

NO

2

 0-10 ppm, NO

2

 0-50 ppm and CI

2

 do not have this feature.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Содержание 5306 500

Страница 1: ...Operating Instructions Compur Monitox plus Subject to changes without notice Copyright Compur Monitors M nchen Editor Compur Monitors GmbH Co KG Weissenseestr 101 D 81539 M nchen...

Страница 2: ...of device certified for group II category 2 Its intended use is the measurement of toxic gases or oxygen Designed with increased safety it is applicable in zone 1 and zone 2 Please observe the followi...

Страница 3: ...ntly high concentrations z Wear the instrument near the breathing zone using the clip or neck chain z The filter cap should be protected from dripping water and exposure to excessive dust z The Monito...

Страница 4: ...ve 2 1 Gas calibration Place Monitox plus gas adapter part 569747 securely on the filter cap Connect span gas NOTE If using HCN Cl2 or other corrosive gases use non porus tubing such as Tygon lined wi...

Страница 5: ...20 9 oxygen Zero can be adjusted with pure Nitrogen GCAL or electronically ICAL 3 Alarm thresholds Compur Monitors recommends to strictly observe the local regulations for the alarm thresholds Open th...

Страница 6: ...t calibration 4 3 Battery replacement Open the instrument Disconnect and remove the battery pack Remove battery pack cover Replace the batteries Observe the proper polarity as is illustrated on the ba...

Страница 7: ...ion data Temperature is out of range This message is displayed for a short time interval Then the Monitox will go on to operate z The instrument operates out of the temperature compensated range Bring...

Страница 8: ...libration Current calibration Intermediate display when the operation mode has been altered and during the self test Self test in process Battery pre alarm Batt and the actual measured value alternate...

Страница 9: ...nder which our products are used are beyond our control Therefore the user must fully test our products and or information to determine suitability for any intended use application condition or situat...

Страница 10: ...10...

Отзывы: