Monitox plus overview  
 

n

 

Filter cap: 

Gas specific protection for the sensor 

o

 

Sensor

   

p

 

Buzzer

 

q

 

LCD-Display: 

r

 

Switch:

 

              

OFF:

    Instrument is off 

 

TEST:    

Selftest 

     

ON:    

Measuring mode. 

s

 

Button Cal: 

Operate with pin. 

t

 

Lithium batteries: 

In a battery pack - replaceable.  

u

 

LED:  

Flashing when alarm threshold is exceeded or  

               in case of an error.  

v

 

Earphone jack

  

                                                                                                            

 
 
 
1. Product description 
 
1.1 Application

 

z

  The 

Compur Monitox plus

 warns the user when a toxic gas is present or oxygen deficiency    

     occurs. 

z

  It should be worn near the breathing zone. 

z

  The display shows the actual gas concentration in ppm, ppb or %/Vol. 

z

  The 

Compur Monitox plus

 is 

not

 designed to monitor process gas streams or permantly   

     high concentrations. 

z

  Wear the instrument near the breathing zone using the clip or neck chain. 

z

  The filter cap should be protected from dripping water and exposure to excessive dust. 

z

  The Monitox comes with a steel clip and neck chain. A suspender and belt clip are optional. 

 

 

 

1.2 Functional test 

Move switch in 

“TEST”

 

position

 : 

Electronics, batteries, 

speaker, LED and Display will be tested.  
Move switch in 

“ON”

 

position

:

 Now you can test the Monitox 

using the optional gas generator. Depending on the gas  
generator model, the black switch pin is operated  
automatically or must be held down manually.  
The gas generator produces a test gas for 10 seconds. 
The test gas concentration is slightly above the alarm 
threshold. After 10 s the gas production stops automatically 
and the green LED of the generator goes off. 
Within these 10 seconds the Monitox should give an alarm. 
 
 
 
 
 
 

 

 

Содержание 5306 500

Страница 1: ...Operating Instructions Compur Monitox plus Subject to changes without notice Copyright Compur Monitors M nchen Editor Compur Monitors GmbH Co KG Weissenseestr 101 D 81539 M nchen...

Страница 2: ...of device certified for group II category 2 Its intended use is the measurement of toxic gases or oxygen Designed with increased safety it is applicable in zone 1 and zone 2 Please observe the followi...

Страница 3: ...ntly high concentrations z Wear the instrument near the breathing zone using the clip or neck chain z The filter cap should be protected from dripping water and exposure to excessive dust z The Monito...

Страница 4: ...ve 2 1 Gas calibration Place Monitox plus gas adapter part 569747 securely on the filter cap Connect span gas NOTE If using HCN Cl2 or other corrosive gases use non porus tubing such as Tygon lined wi...

Страница 5: ...20 9 oxygen Zero can be adjusted with pure Nitrogen GCAL or electronically ICAL 3 Alarm thresholds Compur Monitors recommends to strictly observe the local regulations for the alarm thresholds Open th...

Страница 6: ...t calibration 4 3 Battery replacement Open the instrument Disconnect and remove the battery pack Remove battery pack cover Replace the batteries Observe the proper polarity as is illustrated on the ba...

Страница 7: ...ion data Temperature is out of range This message is displayed for a short time interval Then the Monitox will go on to operate z The instrument operates out of the temperature compensated range Bring...

Страница 8: ...libration Current calibration Intermediate display when the operation mode has been altered and during the self test Self test in process Battery pre alarm Batt and the actual measured value alternate...

Страница 9: ...nder which our products are used are beyond our control Therefore the user must fully test our products and or information to determine suitability for any intended use application condition or situat...

Страница 10: ...10...

Отзывы: