16
a.
Trigger Level = -1kV.
b.
Slope transition “High to Low”
19.
Charge and Trigger the pulse at 5000V.
a.
Push the yellow
CHARGE
button to start charging the internal high voltage capacitor.
b.
Once the desired voltage is reached, press the red TRIGGER button to deliver the pulse (be sure not
to touch the output and return leads when pressing the trigger switch).
20.
Verify the amplitude of the negative pulse captured on the digital oscilloscope:
a.
Vpeak about - 4.72KV (When inductor selected is 500uH).
b.
Vpeak about - 3.88KV (When inductor selected is 25mH).
Figure 4. Checkout Pulse Setup
Figure 5. Vpeak = 4.72kV with 500uH inductor
Figure 6. Vpeak = 3.88kV with 25mH inductor
1000:1
HV Probe