Section 2 

LCB-1010-B Configuration 

 
This manual contains a description of the three configurations that make up your LCB-1010-B instrument.  
Each configuration contains only one leakage current network. 
 
Your  instrument  is  warranted  for  a  period  of  one  year  upon  shipment  of  the  instrument  to  the  original 
purchaser.  

Specifications 

 

Each network configuration includes an input fuse to protect against excessive current that could damage the 
leakage  current  network.    If  the  LCB-1010-B  is  used  incorrectly,  the  internal  fusing  may  not  protect  the 
internal  network  components  from  damage.    Such  damage  is  not  covered  under  warranty.    For  continued 
protection  against  risk  of  fire  or  electric  shock,  replace  the  internal  fuse  only  with  a  type  and  rating  as 
specified in this manual. 

 

Voltage Rating 

0-264Vrms AC to enclosure, O.V. Cat II, 50-440Hz at input terminals

1

 

Current Rating 

100mA maximum 

Fuse Rating (F1) 

Littelfuse 370 series, type TR5, rated 63mA, 250V, fast-acting

2

 

Environmental 

10-90°rH, non-condensing; 0-2000m altitude; 10-40°C 

Components 

High-frequency, low-parasitic, individually selected to maintain 

+/- 1.0% tolerance 

Frequency Response

3

 

50-60 Hz: +/- 3.5% accuracy 

>60 Hz – 1MHz: +/- 5% accuracy 

 

Note 1: Current exceeding the current rating will cause the fuse to open 

 Note 2: Fuse is specified to hold for 1 hour (not open) at 150% of the fuse current rating 
 Note 3: LCB observed accuracy is in accordance with CTL Decision Sheet no. DSH 251b and e. See 
Appendix A for tolerance graphs. 

Configuration 1 

IEC 61010-1:2010 Fig. A.1 

Table 1 

Configuration 1 Specifications 

A, B 

Input terminals to connect to EUT 

Rs 

1500 Ohm 

Cs 

0.22 uF 

Rb 

500 Ohm 

R

2

 

10 kOhm 

C

3

 

0.022 uF 

U

1

, U

2

, COM 

Output terminals for connection to voltmeter 

Содержание LCB-1010-B

Страница 1: ......

Страница 2: ......

Страница 3: ...the art accuracy for making leakage current measurements To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this manual Compliance West USA s...

Страница 4: ......

Страница 5: ...1010 1 2010 Fig A 4 3 Section 3 4 Operating Instructions 4 Setting up your Tester 4 Fuse Verification Check 5 Verification of Proper Operation 5 Section 4 6 Maintenance and Calibration 6 Introduction...

Страница 6: ...y must be used with a suitable voltmeter to perform leakage current measurements The voltmeter used can significantly affect the results of the measurement The voltmeter used must have adequate freque...

Страница 7: ...s specified in this manual Voltage Rating 0 264Vrms AC to enclosure O V Cat II 50 440Hz at input terminals1 Current Rating 100mA maximum Fuse Rating F1 Littelfuse 370 series type TR5 rated 63mA 250V f...

Страница 8: ...o connect to EUT Rs 1500 Ohm Cs 0 22 uF Rb 500 Ohm C D Output terminals for connection to voltmeter Configuration 3 IEC 61010 1 2010 Fig A 4 Table 3 Configuration 3 Specifications A B Input terminals...

Страница 9: ...the front of the instrument One end of each cable is connected to a detachable terminal block as shown in Error Reference source not found below The other end of each cable is unterminated these wires...

Страница 10: ...the following table to know the measurements points and the expected value depending on your LCB configuration If any of your measurements are out of tolerance please contact the manufacturer for fur...

Страница 11: ...anufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the manufacturer General Maintenance To avoid contaminating the PWB with oil from your f...

Страница 12: ...e West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA...

Страница 13: ...000 1500 2000 2500 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 0 100 200 300 400 500 600 10 100 1K 10K 100K 1M Transfer impedance Frequency Hz T...

Страница 14: ...0 1500 2000 2500 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 470 480 490 500 510 520 530 10 100 1K 10K 100K 1M Transfer impedance Frequency Hz T...

Страница 15: ...500 600 700 800 900 1000 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 470 480 490 500 510 520 530 10 100 1K 10K 100K 1M Transfer impedance Frequ...

Отзывы: