background image

 

Tolerance comparison for LCB Configuration 2 (Fig.A.3) 

 

 

Figure A4. 

Input Impedance vs Frequency

 

 

 

Figure A5. 

Transfer Impedance vs Frequency

 

 

 

Figure A6. 

Vout/Vin Ratio vs Frequency 

0

500

1000

1500

2000

2500

10

100

1K

10K

100K

1M

In

p

u

im

p

e

d

an

ce

 [Ω

]

Frequency [Hz]

Input Impedance

tolerance limts per 
CTL DSH 251B

470

480

490

500

510

520

530

10

100

1K

10K

100K

1M

Tr

an

sfe

r i

m

pe

da

nce

 [Ω]

Frequency [Hz]

Transfer Impedance

tolerance limts per 
CTL DSH 251B

0

0.2

0.4

0.6

0.8

1

1.2

10

100

1K

10K

100K

1M

Vo

ut

/V

in

 ra

ti

o

Frequency [Hz]

Vout/Vin (for 
Calibration)

tolerance limts per 
CTL DSH 251B

Содержание LCB-1010-B

Страница 1: ......

Страница 2: ......

Страница 3: ...the art accuracy for making leakage current measurements To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this manual Compliance West USA s...

Страница 4: ......

Страница 5: ...1010 1 2010 Fig A 4 3 Section 3 4 Operating Instructions 4 Setting up your Tester 4 Fuse Verification Check 5 Verification of Proper Operation 5 Section 4 6 Maintenance and Calibration 6 Introduction...

Страница 6: ...y must be used with a suitable voltmeter to perform leakage current measurements The voltmeter used can significantly affect the results of the measurement The voltmeter used must have adequate freque...

Страница 7: ...s specified in this manual Voltage Rating 0 264Vrms AC to enclosure O V Cat II 50 440Hz at input terminals1 Current Rating 100mA maximum Fuse Rating F1 Littelfuse 370 series type TR5 rated 63mA 250V f...

Страница 8: ...o connect to EUT Rs 1500 Ohm Cs 0 22 uF Rb 500 Ohm C D Output terminals for connection to voltmeter Configuration 3 IEC 61010 1 2010 Fig A 4 Table 3 Configuration 3 Specifications A B Input terminals...

Страница 9: ...the front of the instrument One end of each cable is connected to a detachable terminal block as shown in Error Reference source not found below The other end of each cable is unterminated these wires...

Страница 10: ...the following table to know the measurements points and the expected value depending on your LCB configuration If any of your measurements are out of tolerance please contact the manufacturer for fur...

Страница 11: ...anufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the manufacturer General Maintenance To avoid contaminating the PWB with oil from your f...

Страница 12: ...e West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA...

Страница 13: ...000 1500 2000 2500 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 0 100 200 300 400 500 600 10 100 1K 10K 100K 1M Transfer impedance Frequency Hz T...

Страница 14: ...0 1500 2000 2500 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 470 480 490 500 510 520 530 10 100 1K 10K 100K 1M Transfer impedance Frequency Hz T...

Страница 15: ...500 600 700 800 900 1000 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 470 480 490 500 510 520 530 10 100 1K 10K 100K 1M Transfer impedance Frequ...

Отзывы: