background image

 

Section 1 

An Introduction to leakage current testing with the LCB-1010-B 

 
Some amount of leakage current exists in all modern electrical and electronic equipment that connects to an 
AC power source.  Leakage current is always unintentional: either due to capacitive coupling from the AC 
mains circuit to ground (such as exists in  AC motors) or as an unintended side effect of EMI suppression 
components (such as capacitors connected from AC mains to the chassis of the equipment).  The amount of 
leakage current and the measurement methods are governed by various end-product safety standards.  These 
standards often refer to the international standard IEC 60990 - Methods of Measurement of Touch Current 
and Protective Conductor Current and other standards (i.e. IEC 61010) for the design and specification of the 
leakage current measurement network. 
 
Compliance West USA offers several different leakage current networks to perform testing to numerous end-
product standards.  These networks comply with the requirements of the International Committee of Testing 
Laboratories  Operational  Procedure  113  (IECEE  CTL-OP  113  Ed.  1.1).    This  document  outlines 
requirements for design, calibration, and frequency response of the leakage measurement instrument. 
 
The  Compliance  West  USA  LCB  leakage  current  measurement  instruments  are  leakage  current  networks 
only - they must be used with a suitable voltmeter to perform leakage current measurements.  The voltmeter 
used  can  significantly  affect  the  results  of  the  measurement.    The  voltmeter  used  must  have  adequate 
frequency response to measure the AC leakage-current voltages, and have an input impedance of 1 Megaohm 
or greater.  
 

Safety Precautions 

 
Leakage  current  testing  is  inherently  dangerous  because  the  Equipment  Under  Test  (EUT)  is  generally 
ungrounded during testing.  In some safety standards, an additional single-fault condition is introduced during 
leakage current testing, which may expose the test operator to hazardous or even lethal voltages.   
 

Test Personnel 

 
Test operators must be properly trained in the procedures for safety performing leakage current testing.  Metal 
jewelry should not be work during testing.  Operators should not allow others into the testing area, especially 
when tests are being conducted.  Organization is to be stressed.  The operator should keep the test area free 
of unused leads and equipment. 

 

Testing Area 

 
The area used for conducting leakage current tests should be as remote as possible from normal production 
line activities.  Only personnel conducting the test should be allowed in the area, and it should be taped or 
roped off to preclude casual entry by other employees.  In addition, the area should be marked "WARNING 
- HIGH VOLTAGE TESTING" or the equivalent to warn others of the nature of the testing taking place. 
 
The bench being used should be non-conductive, and any exposed metal parts should be tied together and 
grounded. 

Содержание LCB-1010-B

Страница 1: ......

Страница 2: ......

Страница 3: ...the art accuracy for making leakage current measurements To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this manual Compliance West USA s...

Страница 4: ......

Страница 5: ...1010 1 2010 Fig A 4 3 Section 3 4 Operating Instructions 4 Setting up your Tester 4 Fuse Verification Check 5 Verification of Proper Operation 5 Section 4 6 Maintenance and Calibration 6 Introduction...

Страница 6: ...y must be used with a suitable voltmeter to perform leakage current measurements The voltmeter used can significantly affect the results of the measurement The voltmeter used must have adequate freque...

Страница 7: ...s specified in this manual Voltage Rating 0 264Vrms AC to enclosure O V Cat II 50 440Hz at input terminals1 Current Rating 100mA maximum Fuse Rating F1 Littelfuse 370 series type TR5 rated 63mA 250V f...

Страница 8: ...o connect to EUT Rs 1500 Ohm Cs 0 22 uF Rb 500 Ohm C D Output terminals for connection to voltmeter Configuration 3 IEC 61010 1 2010 Fig A 4 Table 3 Configuration 3 Specifications A B Input terminals...

Страница 9: ...the front of the instrument One end of each cable is connected to a detachable terminal block as shown in Error Reference source not found below The other end of each cable is unterminated these wires...

Страница 10: ...the following table to know the measurements points and the expected value depending on your LCB configuration If any of your measurements are out of tolerance please contact the manufacturer for fur...

Страница 11: ...anufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the manufacturer General Maintenance To avoid contaminating the PWB with oil from your f...

Страница 12: ...e West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA...

Страница 13: ...000 1500 2000 2500 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 0 100 200 300 400 500 600 10 100 1K 10K 100K 1M Transfer impedance Frequency Hz T...

Страница 14: ...0 1500 2000 2500 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 470 480 490 500 510 520 530 10 100 1K 10K 100K 1M Transfer impedance Frequency Hz T...

Страница 15: ...500 600 700 800 900 1000 10 100 1K 10K 100K 1M Input impedance Frequency Hz Input Impedance tolerance limts per CTL DSH 251B 470 480 490 500 510 520 530 10 100 1K 10K 100K 1M Transfer impedance Frequ...

Отзывы: