10
IE-LGR-UxxxxM-23
Probe installation and cable routing
- observe the installation instructions,
realize the recommended operating positions, avoid electric-power
distribution systems. See the chapter „
“ for more
information.
The front panel of the device is protected by a transparent transport foil.
Remove it after installing the device.
Device set-up
Connect the appropriate probes and signal leads to the device
.
For more information refer to the chapter
•
Install the COMET Vision
software
into your computer
- is available
for free on the address
•
Run COMET Vision software
Connect the device to the computer
- on the datalogger side, use a USB
cable with USB-C connector.
Setting up the device
- with the aid of the
COMET Vision
program, set up
the device identification, the names of the points to be measured,
the recording mode, the alarms and there signalling. If a PIN code is set on
your SIM card, be sure to set it in the Configuration-Modem-General section
before using the logger for the first time, the default setting is No PIN code.
For more information, refer to the chapter „
Running the device
•
After the set-up
- Disconnect the device from the computer and close
the USB connector with a closing cap. Check that all connectors, including
the antenna connector are properly tightened.
The device may be
operated with a USB cable that is permanently connected to a computer
or to a charger. In this case the ingress protection will be reduced to IP 20.
•
Operating the device by means of keys
- In the main display mode you
can switch over between individual channels and the display mode
Current / minimum / maximum values
by means of the keys. By pressing
a certain key combination, you will enter a menu in which you will be able
to switch the device on or off, to delete the Min/Max values and to disable
the alarm signalling. For more information, refer to the chapter „
the datalogger from the keypad
Charging the batteries
- an internal charger will be activated immediately
after connecting the USB charger or after connecting the datalogger
to the computer. The datalogger contains an intelligent charging circuitry
evaluating the battery condition and internal temperature. Charging time
depends on the current battery discharge status The battery charging
process will be initiated only in the case that the battery voltage is low and
the internal temperature ranges between 0°C and 40 °C. Use
the recommended charger type only. During the charging period the ingress
Содержание U0110G
Страница 5: ...IE LGR UxxxxM 23 5 Dataloggers UxxxxM and UxxxxG U0141M U0141G models...
Страница 61: ...IE LGR UxxxxM 23 61 Dimensions U0110M U0110G U8410M U8410G U0121M U0121G...
Страница 62: ...62 IE LGR UxxxxM 23 U0141M U0141G U3120M U3120G U4440M U4440G...
Страница 63: ...IE LGR UxxxxM 23 63 U3121M U3121G U3631M U3631G...
Страница 64: ...64 IE LGR UxxxxM 23 U0141TM U0141TG U0843M U0843G U5841M U5841G U6841M U6841G U7841M U7841G...