Error 2

 - there is a reading „Err2“ on LCD display.  The  CO

concentration measurement error 

or temperature error occurred.  

 

Error 3 - 

there is a reading „Err3“ on LCD display upper line. Error of internal A/D converter 

appeared  (converter  does  not  respond,  probably  damage  of  A/D  converter).  At  this  state 
device  does  not  measure  temperature  and  relative  humidity.  This  error  does  not  affect  CO

2

 

concentration measurement. It is a serious error, contact distributor of the device. 

 
 

Error 4 - 

there  is  a  reading  „Err4“  on  LCD  display.  It  is  internal  device  error  during 

initialization  of  CO

2

  sensor.  Under  this  condition  device  does  not  measure  concentration  of 

CO

2

. Value read from device is -9999. CO

2

 sensor is probably damaged. It is a serious error, 

contact distributor of the device. This error message is displayed also in case, when the device 
is  connected  to  the  PC  via  the  optional  SP003  communication  cable  and  external  power 
supply is not used. 

 
 

Technical parameters 

 

T8148 

– temperature and CO

2

 concentration transmitter  

 

Output: 

 

 

4 to 20 mA 

 

Power voltage: 

 

9 to 30 V 

 

Power consumption: 

1

 

W during normal operation 

 

 

 

 

max. 4

 

W for 50 ms with 15 s period 

 

Output in case of error: 

<3.8 mA or >24 mA 

 

Temperature: 

Accuracy: 

 

± 0,5 °C  

Range:  

 

0 to +50 ºC 

Resolution: 

 

0,1 °C  

Response time:

 

t90 < 12 min (air flow approximately 0,3 m/s) 

 

 

 

t90 < 25 min (no air flow) 

 

Concentration of CO

2

Accuracy: 

 

± (50 ppm + 2 % of measuring value) at 25°C (77°F)  and 1013 hPa  

Range:  

 

0 to 2000 ppm 

Temp. dependence:   typ. 2 ppm CO

2

 / ºC in the range 0 to 50 ºC (32 to 122°F) 

Long term stability:   typ. 20 ppm / year 
Resolution: 

 

1 ppm 

Response time: 

t90 < 195 s in „SLOW“ measurement mode 

 

 

 

 

t90 < 75 s in „FAST“ measurement mode 

 
 
 

T8248 

– temperature and CO

2

 concentration transmitter  

 

Output: 

 

 

0 to 10 V 

 

Power voltage: 

 

15 to 30 V 

 

Power consumption: 

0,5

 

W during normal operation 

 

 

 

 

max. 3

 

W for 50 ms with 15 s period 

 

Output in case of error: 

<-0.1V or >10.5V 

 

Содержание T8148

Страница 1: ...48 Programmable indoor transmitter of temperature and CO2 concentration with 4 20mA output T8248 Programmable indoor transmitter of temperature and CO2 concentration with 0 10V output Instruction Manu...

Страница 2: ...rary in FAST mode no software filter is used for calculating the output value This fact adds a noise of typ 30ppm which has to be considered in terms of accuracy Of principle measurement is the measur...

Страница 3: ...s heating elements and other heat sources air handlers windows doors into racks and shelves and similar places For buildings with less thermal insulation there is not suitable to place them on externa...

Страница 4: ...t screwdriver beyond the fastening member about 2 cm deep Then slightly move screwdriver in arrow direction it unlock fastening member and the device is partially opened Remove the screwdriver and tak...

Страница 5: ...utput range settings can be verified without a use of the computer by pressing the button on the left side of the device button is placed behind small hole see picture For button pressing use thin ins...

Страница 6: ...ation cable to the PC Installed USB driver detect connected cable and create virtual COM port inside the PC Unscrew four screws of the device lid a remove the lid If device is already installed to mea...

Страница 7: ...se when the device is connected to the PC via the optional SP003 communication cable and external power supply is not used Technical parameters T8148 temperature and CO2 concentration transmitter Outp...

Страница 8: ...C Operating humidity range 5 to 95 RH no condensation Operating pressure range 850 to 1100 hPa Recommended interval of calibration 5 year CO2 2 years temperature Protection IP20 EMC EN 61326 1 EN 5501...

Отзывы: