background image

Messages displayed on the LCD beyond usual operation 

  

If counter is totally full and counter „overflow“ is not enabled, logging stops 
and logger is switched OFF. Message Cntr FULL (counter full) appears on 
the LCD. It appears also if logger is operated with switched OFF display. 
 
New initialization of logger can occur in switching ON the logger 
(immediately after displaying of all LCD segments for checking) e.g. after 
replacement of totally discharged battery for new one. State is indicated by 
INIT reading. It can be displayed for about 12 s. 
If battery voltage drop occurred since last internal clock setting below 
critical limit or battery disconnection for longer period than approximately 
30 s, after display switch ON (during date and time display) all four arrows 
appears as warning to check or set it again from the computer. However all 
logger functions work without limitation. 
If reading BAT is displayed periodically on LCD upper line (for 1 s with 10 
s interval), the end of estimated battery life is coming – however logger 
functions are not limited. Replace battery as soon as possible! 
If reading BAT is displayed permanently, battery voltage is low and logger 
is not possible to switch ON. If logger was switched ON before it, data 
logging is stopped and logger is switched OFF. Communication 
with computer can temporarily work. Replace battery as soon as possible! 
 

 

 

 

  IE-LOG-S7021-04 

Содержание S7021

Страница 1: ...LOGGER S7021 Dual channel logger with counter and binary inputs Instruction for use...

Страница 2: ...aximum value and logger is switched OFF with the message on the LCD If needed counter mode can be selected when after fulfilling counter overflows and continues from value 0 Logging mode can be adjust...

Страница 3: ...me clock adjustable from computer integrated calendar including leap years Error of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download...

Страница 4: ...setting of the instrument accordingly to the user needs menu Configuration Setting of instrument parameters Before logging start it is necessary check or optionally set the logger real time clock sele...

Страница 5: ...e logger LCD indication ON and OFF is displayed permanently Selection of display of increments at the card of counter channel setting causes record of counter increments in time specified by logging i...

Страница 6: ...mperature range battery life can be reduced to 75 i e if displayed remaining battery capacity drops below 25 it is recommended to replace the battery If LCD is switched ON actual readings are displaye...

Страница 7: ...ation decimal point goes out logger would be switched OFF Display goes out automatically after 30 s Remove magnet from slots anytime during display is ON or later Activation of switched OFF LCD for 30...

Страница 8: ...clock setting below critical limit or battery disconnection for longer period than approximately 30 s after display switch ON during date and time display all four arrows appears as warning to check o...

Страница 9: ...s identical with the above procedure for switching ON If decimal point does not appear after 1 s it is necessary remove the magnet and repeat the procedure Logger connection to monitored device by the...

Страница 10: ...l logger settings remain unchanged In the opposite case check by means of PC program all settings especially the real time clock in logger Attention inserted battery with incorrect polarity causes log...

Отзывы: